The SelectSecure catheter delivery system and active fixation 4.1 French diameter Model 3830 pacing lead (Medtronic, Inc, Minneapolis, MN, USA) has the advantage of permitting precise implant at nearly any desired location due to catheter maneuverability. However, as the lead lacks an internal stylet lumen, it is intrinsically floppy and any repositioning after the delivery catheter is removed is nearly impossible, necessitating lead extraction and a repeat of the implant process. At present, there are no commercially available tools to precisely reposition this lead. This report presents a simplified approach to reimplant a dislodged Model 3830 lead using currently available materials to create an effective maneuverable and removable encircling snare.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.