The effects of platelet counts, hematocrit, and leukocyte counts were studied on the closure times of the Thrombostat 4000 (in-vitro bleeding time, IVBT). Closure times became longer with platelet counts <50 x 10'!L; an inverse linear correlation could be established. Hematocrit was also inversely correlated with the closure time. At constant platelet counts a hematocrit of 55% yielded an immediate closure of the filter, while with a hematocrit <15% no closure times could be measured. At constant platelet counts and hematocrits, nomonuclear and polymorphonuclear leukocytes also inDuenced closure times; increased counts resulted in shorter closure times. Leukocytes from a patient with chronic myelocytic leukemia had the same effects.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.