An atrial detector electrode was introduced by mediastinoscopy in 82 patients requiring permanent cardiac pacing. There were no complications. An adequate P wave was recorded in 80 patients. During the following week, the P wave became ineffective in 5 patients; angina occurred in 2 and atrial arrhythmias in 2. Atrially triggered ventricular pacing established in 73 patients and was followed in 71 patients for a period of 1 to 113 months. In 17 cases, it had to be terminated because of an ineffective or unstable P wave,in 6 cases because of atrial arrhythmias, and in 4 cases because of advanced age and recurrent infections. The method is technically simple and place little stress on the patient.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.