During normal rate programming of a pulse generator from slower to more rapid rates, intervals of output impulse suppression lasting 2 seconds plus the newly programmed interval to a maximum of 3.935 seconds are possible, and are a result of the pulse interval control system. These occur if programming falls within the susceptible window (the period between the existing interval and the new one) but will not take place if the programming command is given immediately after an output pulse. Four incidents with resulting ventricular asystole of up to 2.8 seconds are documented and explained. No significant clinical complications were observed.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.