A stability-indicating capillary zone electrophoresis (CZE) method was validated for the analysis of entecavir in pharmaceutical formulations, using nimesulide as an internal standard. A fused-silica capillary (50 μm i.d.; effective length, 40 cm) was used while being maintained at 25 C; the applied voltage was 25 kV. A background electrolyte solution consisted of a 20 mM sodium tetraborate solution at pH 10. Injections were performed using a pressure mode at 50 mbar for 5 s, with detection at 216 nm. The specificity and stability-indicating capability were proven through forced degradation studies, evaluating also the in vitro cytotoxicity test of the degraded products. The method was linear over the concentration range of 1 -200 μg mL -1 (r 2 = 0.9999), and was applied for the analysis of entecavir in tablet dosage forms. The results were correlated to those of validated conventional and fast LC methods, showing non-significant differences (p > 0.05).
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.