Purpose: Verify if the use of thyroid shield was effective in reducing the radiation dose received by the thyroid in mammography. Materials and Methods:Using a General Electric Senograph DMR+ mammographer with two bucky's, thermoluminescent dosimeters and two phantoms, one of body and the other of breast, were made measurements in cranio-caudal and medio-lateral oblique views. The variation was made with and without thyroid shield, and in the breast without shield, on both views. Results:After collecting and analyzing data it was verified a 89% reduction of the entrance skin doses on thyroid with protection. Conclusion:Given the significant reduction of the dose and the principle ALARA we can conclude that the shield should be used by the Radiology service. However the use of the shield may provide artifacts, making difficult the analysis of the breast tissue and leading to the examination repetition increasing the radiation from the mammary gland, as well as the other organs.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations鈥揷itations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.