BackgroundThe software PRIMO for the Monte Carlo simulation of radiotherapy linacs could potentially act as a independent calculation system to verify the calculations of treatment planning systems. We investigated the suitability of the PRIMO default beam parameters to produce accurate dosimetric results for 6 MV photon beams from Varian Clinac 2100 linacs and 6 MV flattening–filter–free photon beams from Varian TrueBeam linacs.MethodsSimulation results with the DPM algorithm were benchmarked against a published reference dosimetry dataset based on point measurements of 25 dosimetric parameters on a large series of linacs. Studied parameters (for several field sizes and depths) were: PDD, off–axis ratios, and output factors for open fields and IMRT/SBRT–style fields. For the latter, the output factors were also determined with radiochromic film and with a small–sized ionization chamber. Benchmark data, PRIMO simulation results and our experimental results were compared.ResultsPDD, off–axis ratios, and open–field output factors obtained from the simulations with the PRIMO default beam parameters agreed with the benchmark data within 2.4% for Clinac 2100, and within 1.3% for TrueBeam. Higher differences were found for IMRT/SBRT–style output factors: up to 2.8% for Clinac 2100, and up to 3.3% for TrueBeam. Experimental output factors agreed with benchmark data within 1.0% (ionization chamber) and within 1.9% (radiochromic film).ConclusionsPRIMO default initial beam parameters for 6 MV photon beams from Varian Clinac 2100 linacs and 6 MV FFF photon beams from Varian TrueBeam linacs allowed agreement within 3.3% with a dosimetry database based on measurements of a high number of linacs. This finding represents a first step in the validation of PRIMO for the independent verification of radiotherapy plans.Electronic supplementary materialThe online version of this article (10.1186/s13014-018-1076-0) contains supplementary material, which is available to authorized users.
Purpose: Conical collimators, or cones, are tertiary collimators that attach to a radiotherapy linac and are suited for the stereotactic radiosurgery treatment of small brain lesions. The small diameter of the most used cones makes difficult the acquisition of the dosimetry data needed for the commissioning of treatment planning systems. Although many publications report dosimetric data of conical collimators for stereotactic radiosurgery, most of the works use different setups, which complicates comparisons. In other cases, the cone output factors reported do not take into account the effect of the small cone diameter on the detector response. Finally, few data exist on the dosimetry of cones with flattening-filter-free (FFF) beams from modern linac models. This work aims at obtaining a dosimetric characterization of the conical collimators manufactured by Brainlab AG (Munich, Germany) in a 6 MV FFF beam from a TrueBeam STx linac (Varian Medical Systems). Methods: Percentage depth dose curves, lateral dose profiles and cone output factors were obtained using Monte Carlo simulations for the cones with diameters of 4,
For the three studied clinical sites, establishing = 400 led to the optimum MU reduction, maintaining the original dose distribution and dosimetric parameters practically unaltered.
The use of the transit image obtained with the electronic portalimaging device (EPID) is becoming an extended method to perform in-vivo dosimetry. The transit images acquired during each fraction can be compared with a predicted image, if available, or with a baseline image, usually the obtained in the first fraction. This work aims to study the dosimetric impact of the failing fractions and to evaluate the appropriateness of using a baseline image in breast plans. Material and methods: Twenty breast patients treated in a Halcyon were retrospectively selected. For each patient and fraction, the treatment plan was calculated over the daily CBCT image. For each fraction, the differences respect to the treatment plan values of OARs and PTV dosimetric parameters were analyzed: ΔD mean , ΔD95%, ΔD98%, ΔD2%, ΔV36Gy, ΔV38.5Gy, and ΔV43.5Gy. Daily fractions were ranked according to the differences found in the dosimetric parameters between the treatment plan and the daily CBCT to establish the best fraction. The daily transit images acquired in every fraction were compared to the first fraction using the global gamma index with the Portal Dosimetry tool. The comparison was repeated using the best fraction image as a baseline.We assessed the correlation of the dosimetric differences obtained from the CBCT images-based treatment plans with the gamma index passing rates obtained using first fraction and best fraction as baseline.Results: Average values of -11.6% [-21.4%, -3.3%] and -3.2% [-1.0%, -10.3%] for the ∆PTVD98% and ∆PTVD95% per every 10% decrease in the passing rate were found, respectively. When using the best fraction as baseline patients were detected with failing fractions that were not detected with the first fraction as baseline.
Conclusion:The gamma passing rates of daily transit images correlate with the coverage loss parameters in breast IMRT plans. Using first fraction image as baseline can lead to the non-detectability of failing fractions.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.