International audienceA new trace glucose analyzer has been designed using electrochemical sensors. The differential device includes (a) a glucose sensor consisting of a modified gas electrode in which ihe pH detector was replaced by a platinum disk and the porous film by a collagen membrane on which beta-D-glucose oxidase has been covalently bound after an acyl-azide activation process; (b) a compensating electrode mounted with a nonenzymatic collagen membrane. After injection of a glucose containing sample into the reaction vessel, where the probes are immersed, an anodic current is detected at the enzyme working electrode. Current outputs of both electrodes are subtracted and twice differentiated; a steady state is reached and the stationary and dynamic responses are recorded. Both responses are proportional to glucose concentration in the 0.1 µM-2 mM range, and the reproducibility was found to be better than 2% using these conditions. The extreme sensitivity exhibited by our system, i.e., 10 nM, is better than previously reported data by 3 orders of magnitude, and is very favorable for trace glucose assays in food and biological samples
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.