The results of our study suggest that stroke volume variation and its surrogate pulse pressure variation derived from pulse contour analysis using an improved algorithm can serve as indicators of fluid responsiveness in normoventilated cardiac surgical patients. Whenever changes in systemic vascular resistance are expected, the PiCCO plus system should be recalibrated.
IntroductionAssessing cardiac preload and fluid responsiveness accurately is important when attempting to avoid unnecessary volume replacement in the critically ill patient, which is associated with increased morbidity and mortality. The present clinical trial was designed to compare the reliability of continuous right ventricular end-diastolic volume (CEDV) index assessment based on rapid response thermistor technique, cardiac filling pressures (central venous pressure [CVP] and pulmonary capillary wedge pressure [PCWP]), and transesophageal echocardiographically derived evaluation of left ventricular end-diastolic area (LVEDA) index in predicting the hemodynamic response to volume replacement.MethodsWe studied 21 patients undergoing elective coronary artery bypass grafting. After induction of anesthesia, hemodynamic parameters were measured simultaneously before (T1) and 12 min after volume replacement (T2) by infusion of 6% hydroxyethyl starch 200/0.5 (7 ml/kg) at a rate of 1 ml/kg per min.ResultsThe volume-induced increase in thermodilution-derived stroke volume index (SVITD) was 10% or greater in 19 patients and under 10% in two. There was a significant correlation between changes in CEDV index and changes in SVITD (r2 = 0.55; P < 0.01), but there were no significant correlations between changes in CVP, PCWP and LVEDA index, and changes in SVITD. The only variable apparently indicating fluid responsiveness was LVEDA index, the baseline value of which was weakly correlated with percentage change in SVITD (r2 = 0.38; P < 0.01).ConclusionAn increased cardiac preload is more reliably reflected by CEDV index than by CVP, PCWP, or LVEDA index in this setting of preoperative cardiac surgery, but CEDV index did not reflect fluid responsiveness. The response of SVITD following fluid administration was better predicted by LVEDA index than by CEDV index, CVP, or PCWP.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.