Multiecho chemical shift-based water-fat separation methods are seeing increasing clinical use due to their ability to estimate and correct for field inhomogeneities. Previous chemical shiftbased water-fat separation methods used a relatively simple signal model that assumes both water and fat have a single resonant frequency. However, it is well known that fat has several spectral peaks. This inaccuracy in the signal model results in two undesired effects. First, water and fat are incompletely separated. Second, methods designed to estimate T* 2 in the presence of fat incorrectly estimate the T* 2 decay in tissues containing fat. In this work, a more accurate multifrequency model of fat is included in the iterative decomposition of water and fat with echo asymmetry and least-squares estimation (IDEAL) water-fat separation and simultaneous T* 2 estimation techniques. The fat spectrum can be assumed to be constant in all subjects and measured a priori using MR spectroscopy. Alternatively, the fat spectrum can be estimated directly from the data using novel spectrum self-calibration algorithms. The improvement in water-fat separation and T* 2 estimation is demonstrated in a variety of in vivo applications, including knee, ankle, spine, breast, and abdominal scans. Key words: water-fat separation; R* 2 measurement; T * 2 measurement; fat spectrum; fat quantification; fat spectral peak Multiecho chemical shift-based water-fat separation methods have seen a recent increase in clinical use (1-6), particularly in challenging applications where inhomogeneous magnetic fields cause failure of conventional fat saturation methods. Dixon (1) first used in-phase (IP) and out-of-phase (OP) images to analytically calculate the water and fat images, in the so-called the 2-point Dixon method. Glover (2) and Glover and Schneider (3) then extended the idea to collect three echoes such that the water-fat separation can be performed with the correction for B 0 field inhomogeneity. In the last decade, numerous variations have been proposed based on the 2-point and 3-point Dixon methods. These previous methods assumed a relatively simple signal representation that models both water and fat as a single resonant frequency. For most applications, this is a satisfactory model and excellent qualitative water-fat separation can be achieved.Although water is well modeled by a single frequency, this is not true for fat. In general, it is assumed that fat resonates at a single frequency ϳ3.5 ppm downfield from water (approximately 210 Hz at 1.5T, and 420 Hz at 3T). However, it is well known that fat has a number of spectral peaks (7-17). In particular, the spectral peak from olefinic proton (5.3 ppm) is close to the water resonant frequency, which will manifest as a baseline level of signal within adipose tissue on the separated water images (2,14,16). This effect is also commonly seen on images acquired with either conventional fat saturation (18) or spatial-spectral excitation (19). In general, this small signal within the fatty tissues is cl...
Quantification of hepatic steatosis is a significant unmet need for the diagnosis and treatment of patients with nonalcoholic fatty liver disease (NAFLD). MRI is capable of separating water and fat signals in order to quantify fatty infiltration of the liver (hepatic steatosis). Unfortunately, fat signal has confounding T 1 effects and the nonzero mean noise in low signal-to-noise ratio (SNR) magnitude images can lead to incorrect estimation of the true lipid percentage. In this study, the effects of bias from T 1 effects and image noise were investigated. An oil/water phantom with volume fat-fractions ranging linearly from 0% to 100% was designed and validated using a spoiled gradient echo (SPGR) sequence in combination with a chemical-shift based fat-water separation method known as iterative decomposition of water and fat with echo asymmetry and least squares estimation (IDEAL). We demonstrated two approaches to reduce the effects of T 1 : small flip angle (flip angle) and dual flip angle methods. Both methods were shown to effectively minimize deviation of the measured fat-fraction from its true value. We also demonstrated two methods to reduce noise bias: magnitude discrimination and phase-constrained reconstruction. Both methods were shown to reduce this noise bias effectively from 15% to less than 1%. Magn Reson Med 58:354 -364, 2007.
Purpose:To describe and demonstrate the feasibility of a novel multiecho reconstruction technique that achieves simultaneous water-fat decomposition and T2* estimation. The method removes interference of water-fat separation with iron-induced T2* effects and therefore has potential for the simultaneous characterization of hepatic steatosis (fatty infiltration) and iron overload. Materials and Methods:The algorithm called "T2*-IDEAL" is based on the IDEAL water-fat decomposition method. A novel "complex field map" construct is used to estimate both R2* (1/T2*) and local B 0 field inhomogeneities using an iterative least-squares estimation method. Water and fat are then decomposed from source images that are corrected for both T2* and B 0 field inhomogeneity. Results:It was found that a six-echo multiecho acquisition using the shortest possible echo times achieves an excellent balance of short scan and reliable R2* measurement. Phantom experiments demonstrate the feasibility with high accuracy in R2* measurement. Promising preliminary in vivo results are also shown. Conclusion:The T2*-IDEAL technique has potential applications in imaging of diffuse liver disease for evaluation of both hepatic steatosis and iron overload in a single breath-hold.
Purpose:To prospectively compare an investigational version of a complex-based chemical shift-based fat fraction magnetic resonance (MR) imaging method with MR spectroscopy for the quantifi cation of hepatic steatosis. Materials and Methods:This study was approved by the institutional review board and was HIPAA compliant. Written informed consent was obtained before all studies. Fifty-fi ve patients (31 women, 24 men; age range, 24-71 years) were prospectively imaged at 1.5 T with quantitative MR imaging and single-voxel MR spectroscopy, each within a single breath hold. The effects of T2* correction, spectral modeling of fat, and magnitude fi tting for eddy current correction on fat quantifi cation with MR imaging were investigated by reconstructing fat fraction images from the same source data with different combinations of error correction. Single-voxel T2-corrected MR spectroscopy was used to measure fat fraction and served as the reference standard. All MR spectroscopy data were postprocessed at a separate institution by an MR physicist who was blinded to MR imaging results. Fat fractions measured with MR imaging and MR spectroscopy were compared statistically to determine the correlation ( r 2 ), and the slope and intercept as measures of agreement between MR imaging and MR spectroscopy fat fraction measurements, to determine whether MR imaging can help quantify fat, and examine the importance of T2* correction, spectral modeling of fat, and eddy current correction. Two-sided t tests (signifi cance level, P = .05) were used to determine whether estimated slopes and intercepts were signifi cantly different from 1.0 and 0.0 , respectively. Sensitivity and specifi city for the classifi cation of clinically signifi cant steatosis were evaluated. Results:Overall, there was excellent correlation between MR imaging and MR spectroscopy for all reconstruction combinations. However, agreement was only achieved when T2* correction, spectral modeling of fat, and magnitude fi tting for eddy current correction were used ( r 2 = 0.99; slope 6 standard deviation = 1.00 6 0.01, P = .77; intercept 6 standard deviation = 0.2% 6 0.1, P = .19 ). Conclusion:T1-independent chemical shift-based water-fat separation MR imaging methods can accurately quantify fat over the entire liver, by using MR spectroscopy as the reference standard, when T2* correction, spectral modeling of fat, and eddy current correction methods are used.q RSNA, 2011
Purpose: To combine gradient-echo (GRE) imaging with a multipoint water-fat separation method known as "iterative decomposition of water and fat with echo asymmetry and least squares estimation" (IDEAL) for uniform waterfat separation. Robust fat suppression is necessary for many GRE imaging applications; unfortunately, uniform fat suppression is challenging in the presence of B 0 inhomogeneities. These challenges are addressed with the IDEAL technique. Materials and Methods:Echo shifts for three-point IDEAL were chosen to optimize noise performance of the water-fat estimation, which is dependent on the relative proportion of water and fat within a voxel. Phantom experiments were performed to validate theoretical SNR predictions. Theoretical echo combinations that maximize noise performance are discussed, and examples of clinical applications at 1.5T and 3.0T are shown. Results:The measured SNR performance validated theoretical predictions and demonstrated improved image quality compared to unoptimized echo combinations. Clinical examples of the liver, breast, heart, knee, and ankle are shown, including the combination of IDEAL with parallel imaging. Excellent water-fat separation was achieved in all cases. The utility of recombining water and fat images into "in-phase," "out-of-phase," and "fat signal fraction" images is also discussed. Conclusion:IDEAL-SPGR provides robust water-fat separation with optimized SNR performance at both 1.5T and 3.0T with multicoil acquisitions and parallel imaging in multiple regions of the body. GRADIENT-ECHO (GRE) imaging is a rapid MRI method that is used for a variety of applications throughout the body. T 1 -weighted (T 1 W) spoiled gradient-echo (SPGR) sequences are of particular importance for postcontrast imaging in many areas of the body, including the abdomen (1) and breast (2). Noncontrast-enhanced T 1 W SPGR imaging is also highly valuable for assessing cartilage morphology (3).Many T 1 W GRE imaging applications require suppression of fat signal. Fat is bright in these sequences and can potentially obscure underlying pathologies, such as tumor or inflammation. Unfortunately, reliable and uniform fat suppression can be challenging in areas of main field (B 0 ) and RF (B 1 ) inhomogeneities. Examples of challenging applications include imaging of the extremities and areas with unfavorable geometry (e.g., the brachial plexus), off-isocenter imaging, and large field of view (FOV) imaging. Other fat-suppression methods, such as short TI inversion recovery (STIR) (4), are incompatible with rapid GRE imaging because of the need for a long inversion time (approximately 200 msec). Spectral-spatial or water-selective pulses can be combined with GRE imaging, but they are lengthy. Although the fat-water discrimination achieved by these pulses is insensitive to B 1 inhomogeneities, they are still sensitive to B 0 inhomogeneities (5).In 1984 Dixon (6) first described "in-and out-ofphase" imaging, a method that acquires two images at different echo times (TEs), thereby exploiting th...
Purpose To determine the precision and accuracy of hepatic fat-fraction measured with a chemical shift-based MRI fat-water separation method, using single-voxel MR spectroscopy (MRS) as a reference standard. Materials and Methods In 42 patients, two repeated measurements were made using a T1-independent, T2∗-corrected chemical shift-based fat-water separation method with multi-peak spectral modeling of fat, and T2-corrected single voxel MR spectroscopy. Precision was assessed through calculation of Bland-Altman plots and concordance correlation intervals. Accuracy was assessed through linear regression between MRI and MRS. Sensitivity and specificity of MRI fat-fractions for diagnosis of steatosis using MRS as a reference standard were also calculated. Results Statistical analysis demonstrated excellent precision of MRI and MRS fat-fractions, indicated by 95% confidence intervals (units of absolute percent) of [−2.66%,2.64%] for single MRI ROI measurements, [−0.81%,0.80%] for averaged MRI ROI, and [−2.70%,2.87%] for single-voxel MRS. Linear regression between MRI and MRS indicated that the MRI method is highly accurate. Sensitivity and specificity for detection of steatosis using averaged MRI ROI were 100% and 94%, respectively. The relationship between hepatic fat-fraction and body mass index was examined. Conclusion Fat-fraction measured with T1-independent T2∗-corrected MRI and multi-peak spectral modeling of fat is a highly precise and accurate method of quantifying hepatic steatosis.
Purpose: To develop a chemical-shift-based imaging method for fat quantification that accounts for the complex spectrum of fat, and to compare this method with MR spectroscopy (MRS). Quantitative noninvasive biomarkers of hepatic steatosis are urgently needed for the diagnosis and management of nonalcoholic fatty liver disease (NAFLD). Materials and Methods:Hepatic steatosis was measured with "fat-fraction" images in 31 patients using a multiecho chemical-shift-based water-fat separation method at 1.5T. Fat-fraction images were reconstructed using a conventional signal model that considers fat as a single peak at -210 Hz relative to water ("single peak" reconstruction). Fat-fraction images were also reconstructed from the same source images using two methods that account for the complex spectrum of fat; precalibrated and self-calibrated "multipeak" reconstruction. Single-voxel MRS that was coregistered with imaging was performed for comparison.Results: Imaging and MRS demonstrated excellent correlation with single peak reconstruction (r 2 ϭ 0.91), precalibrated multipeak reconstruction (r 2 ϭ 0.94), and self-calibrated multipeak reconstruction (r 2 ϭ 0.91). However, precalibrated multipeak reconstruction demonstrated the best agreement with MRS, with a slope statistically equivalent to 1 (0.96 Ϯ 0.04; P ϭ 0.4), compared to self-calibrated multipeak reconstruction (0.83 Ϯ 0.05, P ϭ 0.001) and single-peak reconstruction (0.67 Ϯ 0.04, P Ͻ 0.001). Conclusion:Accurate spectral modeling is necessary for accurate quantification of hepatic steatosis with MRI.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.
hi@scite.ai
10624 S. Eastern Ave., Ste. A-614
Henderson, NV 89052, USA
Copyright © 2024 scite LLC. All rights reserved.
Made with 💙 for researchers
Part of the Research Solutions Family.