Ultra-low field magnetic resonance imaging (ULF MRI) is an effective imaging technique that applies the ultrasensitive detector of superconducting quantum interference device (SQUID) sensor to detect the MR signal at a microtesla field range. In this work, we designed and developed a SQUID-based ULF MRI system with a frequency-adjustable measurement field, the performance of which was characterized via water phantoms. In order to enhance the MR signals, a 500 mT Halbach magnet was used to prepolarize the magnetization of the sample prior to excitation. The signal-to-noise-ratio (SNR) of the spin-echo- (SE-) based pulse sequence can reach up to 70 in a single scan. The images were then reconstructed successfully by using the maximum likelihood expectation maximization (MLEM) algorithm based on the backprojection imaging method. It was demonstrated that an in-plane resolution of 1.8 × 1.8 mm2 can be achieved which indicated the feasibility of SQUID-based MRI at the ULF.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.