Aim: The purpose of the present research was to assess the stability of temporary anchorage devices used in orthodontic treatment. Methodology: Freshly ovine mandibles were cut in blocks. Twenty-seven miniscrews (diameter 1.6 × 8 mm; G2, Dual Top Anchor System, Jeil Medical, Seoul, Korea) were inserted in the blocks and divided in 2 experimental groups: single miniscrew and the innovated design “Seifi Twin Screw (STS)”. Primary stability was evaluated by Periotest “M”® device. Results: Independent t-test showed a significant difference between 2 experimental groups in periotest evaluation (p< 0.05). STS demonstrated higher primary stability due to its mechanical configuration and design. Conclusion: The STS provides higher primary stability and was found to be effective in increased success rate of miniscrew systems from the standpoint of primary stability.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.