Purpose: To demonstrate the feasibility of urgent endovascular treatment of a chronic type A dissection and contained rupture of the false lumen using a noncustomized triple-branched arch endograft, which necessitated reassignment of the branches to the supra-aortic vessels. Case Report:: A 57-year-old patient with a contained rupture of the descending thoracic aorta, in the setting of a chronic type A dissection and a maximum aortic diameter of 85 mm, was converted to endovascular repair after failure of an open surgical approach. A custom-made triple-branched arch endograft designed for another patient was employed, with concomitant occlusion of the false lumen using a Candy Plug occluder. To adjust the graft’s configuration to the patient’s anatomy, the supra-aortic vessels were not assigned to the originally planned branches. The 12-month follow-up angiography demonstrated a satisfactory result. Conclusion: A noncustomized triple-branched arch endograft can be used in an emergency setting to treat chronic type A dissection, reassigning the branches to the supra-aortic vessels as needed.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.