An assessment of the incubators in use at the Kathmandu Maternity Hospital neonatal unit was undertaken; this was followed by a prospective survey of neonatal temperatures on the unit. In the incubator assessment 11 studies were carried out in five incubators. Three of the thermostats in the five incubators did not work at all and those in the other two incubators were more than 3 degrees C inaccurate. All the incubator thermometers gave recordings less than the actual temperature (with a range of error: 1.3-4.4 degrees C). Six out of the 11 babies studied were hypothermic (defined as rectal temperature < 36 degrees C). In the prospective survey of temperatures a high incidence of hypothermia was found on the neonatal unit at the time of first temperature measurement (64 per cent had a rectal temperature of < 36 degrees C). A significant association between admission hypothermia and mortality was noted. Sixteen per cent (10/64) of babies admitted with temperatures < 36 degrees C died within the first week, compared to 0 (0/36) of those admitted with temperatures > or = 36 degrees C.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.