The aim of this work was to test the suitability of a PTW diamond detector for nonreference condition dosimetry in photon beams of different energy (6 and 25 MV) and field size (from 2.6 cm x 2.6 cm to 10 cm x 10 cm). Diamond behavior was compared to that of a Scanditronix p-type silicon diode and a Scanditronix RK ionization chamber. Measurements included output factors (OF). percentage depth doses (PDD) and dose profiles. OFs measured with diamond detector agreed within 1% with those measured with diode and RK chamber. Only at 25 MV, for the smallest field size, RK chamber underestimated OFs due to averaging effects in a pointed shaped beam profile. Agreement was found between PDDs measured with diamond detector and RK chamber for both 6 MV and 25 MV photons and down to 5 cm x 5 cm field size. For the 2.6 cm x 2.6 cm field size, at 25 MV, RK chamber underestimated doses at shallow depth and the difference progressively went to zero in the distal region. PDD curves measured with silicon diode and diamond detector agreed well for the 25 MV beam at all the field sizes. Conversely, the nontissue equivalence of silicon led, for the 6 MV beam, to a slight overestimation of the diode doses in the distal region, at all the field sizes. Penumbra and field width measurements gave values in agreement for all the detectors but with a systematic overestimate by RK measurements. The results obtained confirm that ion chamber is not a suitable detector when high spatial resolution is required. On the other hand, the small differences in the studied parameters, between diamond and silicon systems, do not lead to a significant advantage in the use of diamond detector for routine clinical dosimetry.
The overall uncertainty associated with the use of XR-RV3 films to determine skin dose in the interventional environment can realistically be estimated to be around 20% (k = 1). This uncertainty can be reduced to within 5% if carefully monitoring scanner, film, and fitting-related errors or it can easily increase to over 40% if minimal care is not taken. This work demonstrates the importance of appropriate calibration, reading, fitting, and other film-related and scan-related processes, which will help improve the accuracy of skin dose measurements in interventional procedures.
In this work some dosimetric characteristics of MD-55-2 GafChromic films were studied in a low energy proton beam (21.5 MeV) directly in a water phantom. The nonlinearity of the optical density was quantified by a factor P(lin). A correction factor P(en), that accounts for optical density dependence on the energy, was empirically determined. The effects of detector thickness in depth dose measurements and of the film orientation with respect to beam direction were investigated. The results show that the MD-55-2 films provide dose measurements with the films positioned perpendicularly to the proton beam. A dosimetric formalizm is proposed to determine the dose to water at depth d, with films oriented perpendicularly to the beam axis. This formalism uses a calibration factor of the radiochromic film determined directly on the proton beam at a reference depth in water, and the P(lin) factor, that takes into account the nonlinearity of the calibration curve and the P(en) factor that, in turn takes into account the change of proton beam energy in water. The MD-55-2 films with their high spatial resolution and the quasiwater equivalent material are attractive, positioned perpendicularly along the beam axis, for the absolute dose determination of very small beam sizes and modulated proton beams.
The dosimetric properties of two PTW Riga diamond detectors type 60003 were studied in high-energy photon and electron therapy beam. Properties under study were current-voltage characteristic, polarization effect, time stability of response, dose response, dose-rate dependence, temperature stability, and beam quality dependence of the sensitivity factor. Differences were shown between the two detectors for most of the previous properties. Also, the observed behavior was, to some extent, different from what was reported in the PTW technical specifications. The necessity to characterize each diamond detector individually was addressed.
This paper presents the results of an interlaboratory comparison of retrospective dosimetry using the electron paramagnetic resonance method. The test material used in this exercise was glass coming from the touch screens of smart phones that might be used as fortuitous dosimeters in a large-scale radiological incident. There were 13 participants to whom samples were dispatched, and 11 laboratories reported results. The participants received five calibration samples (0, 0.8, 2, 4, and 10 Gy) and four blindly irradiated samples (0, 0.9, 1.3, and 3.3 Gy). Participants were divided into two groups: for group A (formed by three participants), samples came from a homogeneous batch of glass and were stored in similar setting; for group B (formed by eight participants), samples came from different smart phones and stored in different settings of light and temperature. The calibration curves determined by the participants of group A had a small error and a critical level in the 0.37-0.40-Gy dose range, whereas the curves determined by the participants of group B were more scattered and led to a critical level in the 1.3-3.2-Gy dose range for six participants out of eight. Group A were able to assess the dose within 20 % for the lowest doses (<1.5 Gy) and within 5 % for the highest doses. For group B, only the highest blind dose could be evaluated in a reliable way because of the high critical values involved. The results from group A are encouraging, whereas the results from group B suggest that the influence of environmental conditions and the intervariability of samples coming from different smart phones need to be further investigated. An alongside conclusion is that the protocol was easily transferred to participants making a network of laboratories in case of a mass casualty event potentially feasible.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.