Background: Total laboratory automation (TLA) has been shown to increase laboratory efficiency and quality. However, modular automation is smaller, requires less initial capital, and requires less planning than TLA. We engineered and performed clinical trials on a modular robotic preanalytical workcell for coagulation analysis.
Methods: Timing studies were used to quantify the efficiency of the manual processes and to identify areas in the processing of coagulation specimens where bottlenecks and long waiting periods were encountered. We then designed our modular robotic system to eliminate these bottlenecks. Our robotic modular workcell was engineered to allow a choice of specimen introduction manually, by conveyor, or by mobile robot. Additional timing studies were performed during clinical trials of the robotic system.
Results: Prior to automation, the time required for preanalytical processing time was 18–107 min; after automation, it was 45–50 min. Additional improvements in workcell efficiency could be realized when high quality, prelabeled specimens were introduced into the system.
Conclusion: Compared with manual methods, modular automation provides more predictable variation in specimen processing.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.