Background: The aim of this study was to evaluate the variation in the daily output of a linac Varian Edge TrueBeam over a month, using a Daily QA3 (QA3) and machine performance check (MPC), after Task Group 51protocol and baseline data had been established. Methods and Materials: The daily output data were collected for all photons (6, 10 MV, and 6, 10 FFF) and electrons (6, 9, 12, and 15 MeV) energies. A Bland-Altman-Plot and intraclass correlation coefficients were used to assess agreement and reliability. Results: MPC data showed small random deviations of ≤ 2% and agreed with the QA3 measurements. The maximum mean absolute difference between the QA3 and MPC data was observed with 6 MeV (0.5633), with limits of agreement of (−0.5392–1.6658). Conclusion: These results suggest that MPC data closely track QA3 data and that the two forms are interchangeable
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.