This article reports a calibration procedure for the commercial Anton Paar DMA 512 densimeter usable at pressures up to 40 MPa. The method is based on water as a reference substance and requires knowledge of the characteristics of the cell as functions of temperature when subjected to a vacuum lower than 100 Pa. Comparis on with data from the literature indicates a relative accuracy of 0.1% where the classical calibration method (i.e. assuming no dependence of the densimeter performance upon pressure) led to a relative accuracy of 0.8% for the same substances.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.