An innovative procedure for the development of 3D patient-specific implants with drilling guides for acetabular fracture surgery is presented. By using CT data and 3D surgical planning software, a virtual model of the fractured pelvis was created. During this process the fracture was virtually reduced. Based on the reduced fracture model, patient-specific titanium plates including polyamide drilling guides were designed, 3D printed and milled for intra-operative use. One of the advantages of this procedure is that the personalised plates could be tailored to both the shape of the pelvis and the type of fracture. The optimal screw directions and sizes were predetermined in the 3D model. The virtual plan was translated towards the surgical procedure by using the surgical guides and patient-specific osteosynthesis. Besides the description of the newly developed multi-disciplinary workflow, a clinical case example is presented to demonstrate that this technique is feasible and promising for the operative treatment of complex acetabular fractures.
Key Points
Question
What is the feasibility of using a personalized approach for reconstruction of complex fractures?
Findings
In this case series study, 10 patients with severe acetabular fractures were surgically treated using 3-dimensional (3-D) surgical planning, patient-specific osteosynthesis plates, and drilling guides. These were designed, produced, sterilized and clinically applied within 4 days and were associated with an accurate reconstruction of the fractured articular surface and a good 1-year clinical outcome.
Meaning
These findings suggest that 3-D virtual surgical planning, manufacturing, and clinical application of patient-specific osteosynthesis plates and drilling guides were feasible and yielded good clinical outcomes.
Objectives: In case of surgical removal of oral squamous cell carcinomas, a resection of mandibular bone is frequently part of the treatment. Nowadays, such resections frequently include the application of 3D virtual surgical planning (VSP) and guided surgery techniques. In this paper, current methods for 3D VSP leads for optimisation of the workflow, and patient-specific application of guides and implants are reviewed. Recent findings: Current methods for 3D VSP enable multi-modality fusion of images. This fusion of images is not restricted to a specific software package or workflow. New strategies for 3D VSP in Oral and Maxillofacial Surgery include finite element analysis, deep learning and advanced augmented reality techniques. These strategies aim to improve the treatment in terms of accuracy, predictability and safety. Conclusions: Application of the discussed novel technologies and strategies will improve the accuracy and safety of mandibular resection and reconstruction planning. Accurate, easy-to-use, safe and efficient three-dimensional VSP can be applied for every patient with malignancies needing resection of the mandible.
Over the last decennia, a large variety of osteosynthesis plates and prostheses has been presented in the literature for the treatment of, for example, trauma, oncology or TMJ patients. These applications have in common that failure (loosening or fracturing) of the implanted material, or the screw-fixation of the implant to the patient's bone will often result in failure of the application. It appears crucial for success that the implanted osteosynthesis plates and its fixations are of a matching strength for the patient's specific situation. Mandibular reconstruction failure through either osteosynthesis plate failure or screw loosening are widely reported in the literature as the most common
Background
The aim of this study was to introduce a complete 3D workflow for immediate implant retained prosthetic rehabilitation following maxillectomy in cancer surgery. The workflow consists of a 3D virtual surgical planning for tumor resection, zygomatic implant placement, and for an implant-retained prosthetic-obturator to fit the planned outcome situation for immediate loading.
Materials and methods
In this study, 3D virtual surgical planning and resection of the maxilla, followed by guided placement of 10 zygomatic implants, using custom cutting and drill/placement-guides, was performed on 5 fresh frozen human cadavers. A preoperatively digitally designed and printed obturator prosthesis was placed and connected to the zygomatic implants. The accuracy of the implant positioning was obtained using 3D deviation analysis by merging the pre- and post-operative CT scan datasets.
Results
The preoperatively designed and manufactured obturator prostheses matched accurately the per-operative implant positions. All five obturators could be placed and fixated for immediate loading. The mean prosthetic point deviation on the cadavers was 1.03 ± 0.85 mm; the mean entry point deviation was 1.20 ± 0.62 mm; and the 3D angle deviation was 2.97 ± 1.44°.
Conclusions
It is possible to 3D plan and accurately execute the ablative surgery, placement of zygomatic implants, and immediate placement of an implant-retained obturator prosthesis with 3D virtual surgical planning.The next step is to apply the workflow in the operating room in patients planned for maxillectomy.
Zygomatic implants are used in patients with maxillary defects to improve the retention and stability of obturator prostheses, thereby securing good oral function. Prosthetic-driven placement of zygomatic implants is even difficult for experienced surgeons, and with a free-hand approach, deviation from the preplanned implant positions is inevitable, thereby impeding immediate implant-retained obturation. A novel, digitalized workflow of surgical planning was used in 10 patients. Maxillectomy was performed with 3D-printed cutting, and drill guides were used for subsequent placement of zygomatic implants with immediate placement of implant-retained obturator prosthesis. The outcome parameters were the accuracy of implant positioning and the prosthetic fit of the obturator prosthesis in this one-stage procedure. Zygomatic implants (n = 28) were placed with good accuracy (mean deviation 1.73 ± 0.57 mm and 2.97 ± 1.38° 3D angle deviation), and in all cases, the obturator prosthesis fitted as pre-operatively planned. The 3D accuracy of the abutment positions was 1.58 ± 1.66 mm. The accuracy of the abutment position in the occlusal plane was 2.21 ± 1.33 mm, with a height accuracy of 1.32 ± 1.57 mm. This feasibility study shows that the application of these novel designed 3D-printed surgical guides results in predictable zygomatic implant placement and provides the possibility of immediate prosthetic rehabilitation in head and neck oncology patients after maxillectomy.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.