SummaryWe prospectively assessed common clinical endpoints for their usefulness in avoiding hyperinflation of the cuffs of laryngeal mask airways (slight outward movement) and tracheal tubes (disappearance of an audible leak around the cuff during manual ventilation < 20 cmH 2 O) in 640 children. Cuff pressures were measured at induction and immediately before emergence from anaesthesia. With the laryngeal mask airway (sizes 1-4), the median cuff pressures ranged from 90 to > 120 cmH 2 O at induction and 105 to > 120 cmH 2 O before emergence. With tracheal tubes (sizes 3-7 mm), median cuff pressures were 40-60 cmH 2 O at induction and 45-70 cmH 2 O at emergence. With the use of nitrous oxide a consistent rise in cuff pressure was observed between the first and second readings whereas cuff pressures remained constant when nitrous oxide was not used. The use of clinical endpoints alone was associated with significant hyperinflation of cuffs with both devices in almost all patients, with an exacerbation when nitrous oxide was used. In order to avoid unnecessary cuff hyperinflation in laryngeal mask airways and tracheal tubes, the routine use of cuff manometers is mandatory in children.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.