The DS18B20 sensor calibration has been conducted using a fluke hart with a temperature chamber at the center for meteorology, climatology, and geophysics region III Denpasar. The calibration process is done by stabilizing the temperature in the chamber, then the temperature on the DS18B20 sensor and the standard thermometer in the chamber are read and compared. The setpoints used were 10 oC, 20 oC, 30 oC and 40 oC each set point was read four times. The results of the calculation of the standard deviation at each setpoint are 0,39 oC, 0,12 oC, 0,00 oC, 0,00 oC and the results of the uncertainty for each setpoint are 0,65 oC, 0,20 oC, 0,18 oC, 0,18 oC. The results of the DS18B20 sensor calibration using a fluke hart with a temperature chamber obtained the highest standard deviation value of 0,39 oC and the highest uncertainty value was 0,65 oC. This value indicates that the DS18B20 sensor calibrated with a standard tool at the 10 oC set point has not met the standard value applied by WMO, which is ±0.3 oC.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.