Background
The use of a functional electromechanical dynamometer (FEMD) has been proposed as a valid and effective tool to evaluate specific movement patterns. The aim of this study was to determine the reliability of FEMD on swing eccentric hamstring exercise (SEHE) measures in soccer players.
Methods
Nineteen federated male soccer players (20.74 ± 4.04 years) performed the SEHE at three different isokinetic velocities (20–40–60 cm/s). These evaluations were conducted in four sessions, two for familiarization and two for registration. The average and maximum load (N) of the three isokinetic velocities was calculated from the values obtained from the FEMD (Dynasystem®, Bangalore).
Results
The main results of this research showed that the reliability was high for the average load in the condition of 40 cm/s, presenting the highest ICC value (0.94). For maximum load, reliability was high in the condition of 20 cm/s. The manifestation of the most reliable load was the maximum load (ICC = 0.91–0.87).
Conclusions
FEMD (Dynasystem®, Bangalore) is a reliable device to evaluate the eccentric strength of the hamstring muscles in soccer players.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.