Purpose:We evaluated the performance of organ contour propagation from a planning computed tomography to cone-beam computed tomography with deformable image registration by comparing contours to manual contouring.Materials and Methods:Sixteen patients were retrospectively identified based on showing considerable physical change throughout the course of treatment. Multiple organs in the 3 regions (head and neck, prostate, and pancreas) were evaluated. A cone-beam computed tomography from the end of treatment was registered to the planning computed tomography using rigid registration, followed by deformable image registration. The contours were copied on cone-beam computed tomography image sets using rigid registration and modified by 2 radiation oncologists. Contours were compared using Dice similarity coefficient, mean surface distance, and Hausdorff distance.Results:The mean physician-to-physician Dice similarity coefficient for all organs was 0.90. When compared to each physician’s contours, the overall mean for rigid was 0.76 (P < .001), and it was improved to 0.79 (P < .001) for deformable image registration. Comparing deformable image registration to physicians resulted in a mean Dice similarity coefficient of 0.77, 0.74, and 0.84 for head and neck, prostate, and pancreas groups, respectively; whereas, the physician-to-physician mean agreement for these sites was 0.87, 0.90, and 0.93 (P < .001, for all sites). The mean surface distance for physician-to-physician contours was 1.01 mm, compared to 2.58 mm for rigid-to-physician contours and 2.24 mm for deformable image registration-to-physician contours. The mean physician-to-physician Hausdorff distance was 11.32 mm, and when compared to any physician’s contours, the mean for rigid and deformable image registration was 12.1 mm and 12.0 mm (P < .001), respectively.Conclusion:The physicians had a high level of agreement via the 3 metrics; however, deformable image registration fell short of this level of agreement. The automatic workflows using deformable image registration to deform contours to cone-beam computed tomography to evaluate the changes during treatment should be used with caution.
Among the various parameters that were used to evaluate the SBRT plans, the CI100% and CI50% were the most challenging criteria to meet. Although the CSRs of organs at risk were higher among all cases, their proximity to the PTV was a significant factor.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.