Advances in radiation treatment with beamlet-based intensity modulation, image-guided radiation therapy, and stereotactic radiosurgery (including specialized equipments like CyberKnife, Gamma Knife, tomotherapy, and high-resolution multileaf collimating systems) have resulted in the use of reduced treatment fields to a subcentimeter scale. Compared to the traditional radiotherapy with fields > or =4 x 4 cm2, this can result in significant uncertainty in the accuracy of clinical dosimetry. The dosimetry of small fields is challenging due to nonequilibrium conditions created as a consequence of the secondary electron track lengths and the source size projected through the collimating system that are comparable to the treatment field size. It is further complicated by the prolonged electron tracks in the presence of low-density inhomogeneities. Also, radiation detectors introduced into such fields usually perturb the level of disequilibrium. Hence, the dosimetric accuracy previously achieved for standard radiotherapy applications is at risk for both absolute and relative dose determination. This article summarizes the present knowledge and gives an insight into the future procedures to handle the nonequilibrium radiation dosimetry problems. It is anticipated that new miniature detectors with controlled perturbations and corrections will be available to meet the demand for accurate measurements. It is also expected that the Monte Carlo techniques will increasingly be used in assessing the accuracy, verification, and calculation of dose, and will aid perturbation calculations of detectors used in small and highly conformal radiation beams. rican Association of Physicists in Medicine.
A method for photon beam dose calculations is described. The primary photon beam is raytraced through the patient, and the distribution of total radiant energy released into the patient is calculated. Polyenergetic energy deposition kernels are calculated from the spectrum of the beam, using a database of monoenergetic kernels. It is shown that the polyenergetic kernels can be analytically described with high precision by (A exp( -ar) + B exp( -br)/r2, where A, a, B, and b depend on the angle with respect to the impinging photons and the accelerating potential, and r is the radial distance. Numerical values of A, a, B, and b are derived and used to convolve energy deposition kernels with the total energy released per unit mass (TERMA) to yield dose distributions. The convolution is facilitated by the introduction of the collapsed cone approximation. In this approximation, all energy released into coaxial cones of equal solid angle, from volume elements on the cone axis, is rectilinearly transported, attenuated, and deposited in elements on the axis. Scaling of the kernels is implicitly done during the convolution procedure to fully account for inhomogeneities present in the irradiated volume. The number of computational operations needed to compute the dose with the method is proportional to the number of calculation points. The method is tested for five accelerating potentials; 4, 6, 10, 15, and 24 MV, and applied to two geometries; one is a stack of slabs of tissue media, and the other is a mediastinum-like phantom of cork and water. In these geometries, the EGS4 Monte Carlo system has been used to generate reference dose distributions with which the dose computed with the collapsed cone convolution method is compared. Generally, the agreement between the methods is excellent. Deviations are observed in situations of lateral charged particle disequilibrium in low density media, however, but the result is superior compared to that of the generalized Batho method.
For commissioning a linear accelerator for clinical use, medical physicists are faced with many challenges including the need for precision, a variety of testing methods, data validation, the lack of standards, and time constraints. Since commissioning beam data are treated as a reference and ultimately used by treatment planning systems, it is vitally important that the collected data are of the highest quality to avoid dosimetric and patient treatment errors that may subsequently lead to a poor radiation outcome. Beam data commissioning should be performed with appropriate knowledge and proper tools and should be independent of the person collecting the data. To achieve this goal, Task Group 106 (TG-106) of the Therapy Physics Committee of the American Association of Physicists in Medicine was formed to review the practical aspects as well as the physics of linear accelerator commissioning. The report provides guidelines and recommendations on the proper selection of phantoms and detectors, setting up of a phantom for data acquisition (both scanning and no-scanning data), procedures for acquiring specific photon and electron beam parameters and methods to reduce measurement errors (<1%), beam data processing and detector size convolution for accurate profiles. The TG-106 also provides a brief.discussion on the emerging trend in Monte Carlo simulation techniques in photon and electron beam commissioning. The procedures described in this report should assist a qualified medical physicist in either measuring a complete set of beam data, or in verifying a subset of data before initial use or for periodic quality assurance measurements. By combining practical experience with theoretical discussion, this document sets a new standard for beam data commissioning.
Dose calculation methods for photon beams are reviewed in the context of radiation therapy treatment planning. Following introductory summaries on photon beam characteristics and clinical requirements on dose calculations, calculation methods are described in order of increasing explicitness of particle transport. The simplest are dose ratio factorizations limited to point dose estimates useful for checking other more general, but also more complex, approaches. Some methods incorporate detailed modelling of scatter dose through differentiation of measured data combined with various integration techniques. State-of-the-art methods based on point or pencil kernels, which are derived through Monte Carlo simulations, to characterize secondary particle transport are presented in some detail. Explicit particle transport methods, such as Monte Carlo, are briefly summarized. The extensive literature on beam characterization and handling of treatment head scatter is reviewed in the context of providing phase space data for kernel based and/or direct Monte Carlo dose calculations. Finally, a brief overview of inverse methods for optimization and dose reconstruction is provided.
A method for photon dose calculation in radio therapy planning using pencil beam energy deposition kernels is presented. It is designed to meet the requirements of an algorithm for 3-D treatment planning that is general enough to handle irregularly shaped radiation fields incident on a heterogeneous patient. It is point oriented and thus faster than a full 3-D convolution algorithm and uses the same physical data base to characterize a clinical beam as a full 3-D convolution algorithm. It is shown that photon therapy beams can be characterized with great accuracy from a combination of precalculated Monte Carlo energy deposition kernels and dose distributions measured in a water phantom. The data are used to derive analytical pencil beam kernels that are approximately partitionated into the dose from (i) primary released electrons and positrons, (ii) scattered, bremsstrahlung, and annihilation photons, (iii) contaminating photons, and (iv) charged particles from the collimator head. A semianalytical integration method, based on triangulation of the field, is developed for dose calculation using the analytical kernels. Dose is calculated in units normalized to the incident energy fluence which facilitates output factor calculation. For application in heterogeneous media, a scatter correction factor is derived using monodirectional convolution along the ray path. In homogeneous media results are compared with measurements and in heterogeneous media with Monte Carlo calculations and the Batho method.
The concept of in-air output ratio (Sc) was introduced to characterize how the incident photon fluence per monitor unit (or unit time for a Co-60 unit) varies with collimator settings. However, there has been much confusion regarding the measurement technique to be used that has prevented the accurate and consistent determination of Sc. The main thrust of the report is to devise a theoretical and measurement formalism that ensures interinstitutional consistency of Sc. The in-air output ratio, Sc, is defined as the ratio of primary collision water kerma in free-space, Kp, per monitor unit between an arbitrary collimator setting and the reference collimator setting at the same location. Miniphantoms with sufficient lateral and longitudinal thicknesses to eliminate electron contamination and maintain transient electron equilibrium are recommended for the measurement of Sc. The authors present a correction formalism to extrapolate the correct Sc from the measured values using high-Z miniphantom. Miniphantoms made of high-Z material are used to measure Sc for small fields (e.g., IMRT or stereotactic radiosurgery). This report presents a review of the components of Sc, including headscatter, source-obscuring, and monitor-backscattering effects. A review of calculation methods (Monte Carlo and empirical) used to calculate Sc for arbitrary shaped fields is presented. The authors discussed the use of Sc in photon dose calculation algorithms, in particular, monitor unit calculation. Finally, a summary of Sc data (from RPC and other institutions) is included for QA purposes.
Abbreviations: %dd(10) x , The photon component of the percent depth dose at 10 cm depth in water for a 10 cm 2 × 10 cm 2 field; L∕ w air , Restricted mass collision stopping power ratio of water to air; en ∕ , Spectrum-averaged mass energy-absorption coefficient; AAPM, American Association of Physicists in Medicine; CPE, Charged Particle Equilibrium; DLG, Dosimetric leaf gap; D f msr w,Q msr , Absorbed dose to water at the reference depth z ref in water in the absence of the detector in a field specified by f msr and beam quality Q msr .; f clin , Clinical (clin) non-reference radiation field; f msr , Machine-specific reference (msr) field; f ref , Reference field (ref) specified in dosimetry protocols for which the calibration coefficient of an ionization chamber in terms of absorbed dose to water is provided by a standards laboratory; FWHM, Full-width at half-maximum; GUM, Guide to the expression of Uncertainty in Measurements.; IAEA, International Atomic Energy Agency; ICRU, International Commission on Radiation Units and Measurements; IMRT, Intensity-modulated radiation therapy; k Q,Q 0 , The beam-quality correction factor, which corrects for the differences between the response of an ionization chamber in the reference beam of quality Q o used for calibrating the chamber and the beam of quality Q (defined as k Q . in TG-51 4 ); k f ref Q,Q 0 , Correction factor that accounts for the differences between the response of a detector in field f ref in a beam of quality Q and reference beam quality Q 0 as defined in TRS-483. 1 and Palmans et al 2 ; k fclin,fmsr Qclin,Q msr , The detector-specific correction factor that accounts for the difference between the responses of the detector in fields f clin in a beam of quality Q clin and in fields f msr in beam of quality Q msr as defined by Alfonso et al. 3 ; LCPE, Lateral charged particle equilibrium; M fmsr Qmsr , Detector reading in field f msr and beam quality Q msr corrected for influence of changes in pressure and temperature, incomplete charge collection, polarity effect and electrometer correction factor (TRS-483 1 ); MU, Monitor unit; N D,w,Q 0 , This is N D,w in TG-51, 4 and defined as the calibration coefficient in terms of absorbed dose to water for an ionization chamber at a reference beam of quality,Q 0 and field size f ref ; N f ref D,w,Q 0
Analytic models for calculation of scatter distributions from flattening filters in therapy photon beams are presented. It is shown that the amount of scatter with high atomic number filters can vary from 2% in 4-MV beams to 10% for 24-MV beams. The use of low atomic number filters can increase the amount of scatter by a factor of 2. The dependence on the opening angle of the primary collimator is quite large since a larger opening angle requires a thicker filter, which increases the scattered fraction of the filtered beam. The scatter makes the filter act as an extended source of extra-focal radiation. The source distribution for monomedia filters is shown to be almost triangular. Integration in the calculation-point's eye view over the visible part of the filter scatter source yields the scatter fraction of the total energy fluence incident upon the patient. The experimentally well-known "tilt" of dose profiles for asymmetrical fields is explained by the model. For complete modeling of head scatter distributions in treatment planning, the model presented must be combined with models also describing the scatter from the collimators, auxiliary modulators such as wedges and compensating filters, and collimator backscatter to the beam monitor.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.
hi@scite.ai
10624 S. Eastern Ave., Ste. A-614
Henderson, NV 89052, USA
Copyright © 2024 scite LLC. All rights reserved.
Made with 💙 for researchers
Part of the Research Solutions Family.