In this paper, we propose an ellipsometer using a phase retarder and rotating polarizer and analyzer at a speed ratio 1:N . Different ellipsometric configurations are presented by assuming N = 1, 2, and 3. Moreover, two values of the offset angle of the retarder are considered for each ellipsometric configuration. The Mueller formalism is employed to extract the Stokes parameters, from which the intensity received by the detector is obtained. The optical properties of c-Si are calculated using all configurations. A comparison between different configurations is carried out considering the effect of the noise on the results and the uncertainties in the ellipsometric parameters as functions of the uncertainties of the Fourier coefficients. It is found that the alignment of the phase retarder has a crucial impact on the results and the ellipsometric configuration with speed ratio 1:1 is preferred over the other configurations.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.