Background: Reliable point-of-care (POC) diagnostics not requiring laboratory infrastructure could be a game changer in the COVID-19 pandemic, particularly in the Global South. We assessed performance, limit of detection (LOD) and ease-of-use of three antigen-detecting, rapid POC diagnostics (Ag-RDT) for SARS-CoV-2.
Methods: This prospective, multi-centre diagnostic accuracy study, recruited participants suspected to have SARS-CoV2 in Germany and UK. Paired nasopharyngeal swabs (NP) or NP and/or oropharyngeal swabs (OP) were collected from participants (one for clinical real-time reverse transcription polymerase chain reaction (RT-PCR) and one for Ag-RDT testing). Performance of each of three Ag-RDTs was compared to RT-PCR overall, and according to predefined subcategories e.g. cycle threshold (CT)-value, days from symptom onset, etc. In addition, limited verification of analytical limit-of-detection (LOD) was determined. To understand the usability of each Ag-RDT a System Usability Scale (SUS) questionnaire and ease-of-use assessment were performed.
Results: Between April 17th and August 25th, 2020, 2417 participants were enrolled, with 70 (3.0%) testing positive by RT-PCR. The best-performing test (SD Biosensor, Inc. STANDARD Q) was 76.6% [95% Confidence Interval (CI) 62.8-86.4] sensitive and 99.3% [CI 98.6-99.6] specific. A sub-analysis showed all samples with RT-PCR CT-values <25 were detectable by STANDARD Q. The test was considered easy-to-use (SUS 86/100) and suitable for POC. Bioeasy and Coris showed specificity of 93.1% [CI 91.0%-94.8%] and 95.8% [CI 93.4%-97.4%], respectively, not meeting the predefined target of ≥98%.
Conclusion: There is large variability in performance of Ag-RDT tests with one test showing promise. Given the usability at POC, these tests are likely to have impact despite imperfect sensitivity; however further research and modelling are needed.
In the context of the coronavirus disease 2019 (COVID-19) pandemic there has been an increase of the use of antigen-detection rapid diagnostic tests (Ag-RDT). The performance of Ag-RDT vary greatly between manufacturers and evaluating their analytical limit of detection (LOD) has become high priority. Here we describe a manufacturer-independent evaluation of the LOD of 19 marketed Ag-RDT using live SARS-CoV-2 spiked in different matrices: direct culture supernatant, a dry swab, and a swab in Amies. Additionally, the LOD using dry swab was investigated after 7 days’ storage at − 80 °C of the SARS-CoV-2 serial dilutions. An LOD of ≈ 5.0 × 102 pfu/ml (1.0 × 106 genome copies/ml) in culture media is defined as acceptable by the World Health Organization. Fourteen of 19 Ag-RDTs (ActiveXpress, Espline, Excalibur, Innova, Joysbio, Mologic, NowCheck, Orient, PanBio, RespiStrip, Roche, Standard-F, Standard-Q and Sure-Status) exceeded this performance criteria using direct culture supernatant applied to the Ag-RDT. Six Ag-RDT were not compatible with Amies media and a decreased sensitivity of 2 to 20-fold was observed for eleven tests on the stored dilutions at − 80 °C for 7 days. Here, we provide analytical sensitivity data to guide appropriate test and sample type selection for use and for future Ag-RDT evaluations.
PCR of upper respiratory specimens is the diagnostic standard for severe acute respiratory syndrome coronavirus 2 infection. However, saliva sampling is an easy alternative to nasal and throat swabbing. We found similar viral loads in saliva samples and in nasal and throat swab samples from 110 patients with coronavirus disease.
RT-qPCR utilising upper respiratory swabs are the diagnostic gold standard for SARS-CoV-2 despite reported low sensitivity and limited scale up due to global shortages. Saliva is a non-invasive, equipment independent alternative to swabs.
We collected 145 paired saliva and nasal/throat (NT) swabs at diagnosis (day 0) and repeated on day 2 and day 7 dependent on inpatient care and day 28 for study follow up. Laboratory cultured virus was used to determine the analytical sensitivity of spiked saliva and swabs containing amies preservation media.
Self-collected saliva samples were found to be consistent, and in some cases superior when compared to healthcare worker collected NT swabs from COVID-19 suspected participants. We report for the first time the analytical limit of detection of 10-2 and 100 pfu/ml for saliva and swabs respectively.
Saliva is a easily self-collected, highly sensitive specimen for the detection of SARS-CoV-2.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.