IMPORTANCE
Trends and in-hospital outcomes associated with early adoption of the subcutaneous implantable cardioverter defibrillator (S-ICD) in the United States have not been described.
OBJECTIVES
To describe early use of the S-ICD in the United States and to compare in-hospital outcomes among patients undergoing S-ICD vs transvenous (TV)-ICD implantation.
DESIGN, SETTING, AND PARTICIPANTS
A retrospective analysis of 393 734 ICD implants reported to the National Cardiovascular Data Registry ICD Registry, a nationally representative US ICD registry, between September 28, 2012 (US Food and Drug Administration S-ICD approval date), and March 31, 2015, was conducted. A 1:1:1 propensity-matched analysis of 5760 patients was performed to compare in-hospital outcomes among patients with S-ICD with those of patients with single-chamber (SC)–ICD and dual-chamber (DC)–ICD.
MAIN OUTCOMES AND MEASURES
Analysis of trends in S-ICD adoption as a function of total ICD implants and comparison of in-hospital outcomes (death, complications, and defibrillation threshold [DFT] testing) among S-ICD and TV-ICD recipients.
RESULTS
Of the 393 734 ICD implants evaluated during the study period, 3717 were S-ICDs (0.9%). A total of 109 445 (27.8%) of the patients were female; the mean (SD) age was 67.03 (13.10) years. Use of ICDs increased from 0.2%during the fourth quarter of 2012 to 1.9% during the first quarter of 2015. Compared with SC-ICD and DC-ICD recipients, those with S-ICDs were more often younger, female, black, undergoing dialysis, and had experienced prior cardiac arrest. Among 2791 patients with S-ICD who underwent DFT testing, 2588 (92.7%), 2629 (94.2%), 2635 (94.4%), and 2784 (99.7%) were successfully defibrillated (≤65, ≤70, ≤75, and≤80 J, respectively). In the propensity-matched analysis of 5760 patients, in-hospital complication rates associated with S-ICDs (0.9%) were comparable to those of SC-ICDs (0.6%) (P = .27) and DC-ICD rates (1.5%) (P = .11). Mean (SD) length of stay after S-ICD implantation was comparable to that after SC-ICD implantation (1.1 [1.5] vs 1.0 [1.2] days; P = .77) and less than after DC-ICD implantation (1.1 [1.5] vs 1.2 [1.5] days; P < .001).
CONCLUSIONS AND RELEVANCE
The use of S-ICDs is rapidly increasing in the United States. Early adoption has been associated with low complication rates and high rates of successful DFT testing despite frequent use in patients with a high number of comorbidities.