Twenty-three multiprogrammable pacemakers and four implantable cardioverter defibrillators (ICDs) containing either complementary metal-oxide semiconductor (CMOS) or CMOS/Bipolar integrated circuit (IC) technology were exposed to 6-MV photon and 18-MeV electron radiation at various dose levels. Of the 17 pacemakers exposed to photon radiation eight failed before 50 Gy, whereas four of the six pacemakers exposed to electron radiation failed before 70 Gy. Photon scatter doses were well tolerated. For the ICDs detection and charging time increased with accumulated radiation dose, the charging time increased catastrophically at less than 50 total pulses delivered when compared with the charging time of six implanted ICDs. Sensitivity and output energy delivered by the ICD pulse were constant during the test. It was found that devices using the shorter channel length IC technology (i.e., 3 microns CMOS) were per se harder to ionizing radiation than the devices using larger channel length IC technologies (i.e., either 8 microns CMOS or combined 5 microM CMOS/20 V Bipolar). In fact, none of the devices based on 3 microns CMOS IC technology failed before 76 Gy, which is above the highest dose level (70 Gy) normally used in radiation oncology treatments.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.