Purpose
To describe a rare complication of phacoemulsification (Phaco) cataract extraction.
Observations
A phaco tip fracture occurred during the sculpting of the nucleus in a cataract extraction surgery of a 60-year-old woman. The surgeon introduced the second instrument through the side port to retrieve the broken portion of the tip while maintaining it inside the sleeve.
Conclusions and improtance
Phaco tip fracture is a rare but serious complication that all cataract surgeons should be able to manage safely. Main risk factors for phaco tip fractures are: Aspiration Bypass System (ABS) phaco tip design and sterilization and reuse of instruments. A step by step plan of action was described in detail for a surgeon to manage a phaco tip fracture which mainly consists of removing the fractured phaco tip while it is still inside the sleeve to prevent any touch between the sharp broken edge of the tip and any intraocular component.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.