Thank you for choosing to publish with us. This is your final opportunity to ensure your article will be accurate at publication. Please review your proof carefully and respond to the queries using the circled tools in the image below, which are available by clicking "Comment" from the right-side menu in Adobe Reader DC.* Please use only the tools circled in the image, as edits via other tools/methods can be lost during file conversion. For comments, questions, or formatting requests, please use. Please do not use comment bubbles/sticky notes. Comment *If you do not see these tools, please ensure you have opened this file with Adobe Reader DC, available for free at get.adobe.com/reader or by going to Help > Check for Updates within other versions of Reader. For more detailed instructions, please see us.sagepub.com/ReaderXProofs.
Background
Some myocardial diseases, such as cardiac sarcoidosis, predispose to complete atrioventricular block. The European Society of Cardiology Guidelines on cardiac pacing in 2021 recommend myocardial disease screening in patients with conduction disorder requiring pacemaker with multimodality imaging, including cardiac magnetic resonance (CMR) imaging. The ability of CMR imaging to detect myocardial disease in patients with a temporary pacing wire is not well documented.
Methods and Results
Our myocardial disease screening protocol is based on using an active fixation pacing lead connected to a reusable extracorporeal pacing generator (temporary permanent pacemaker) as a bridge to a permanent pacemaker. From 2011 to 2019, we identified 17 patients from our CMR database who underwent CMR imaging with a temporary permanent pacemaker for atrioventricular block. We analyzed their clinical presentations, CMR data, and pacemaker therapy. All CMRs were performed without adverse events. Pacing leads induced minor artifacts to the septal myocardial segments. The extent of late gadolinium enhancement in CMR imaging was used to screen patients for the presence of myocardial disease. Patients with evidence of late gadolinium enhancement underwent endomyocardial biopsy. If considered clinically indicated, also 18‐F‐fluorodeoxyglucose positron emission tomography and extracardiac tissue biopsy were performed if sarcoidosis was suspected. Eventually, 8 of 17 patients (47.1%) were diagnosed with histologically confirmed granulomatous inflammatory cardiac disease. Importantly, only 1 had a previously diagnosed extracardiac sarcoidosis at the time of presentation with high‐degree atrioventricular block.
Conclusions
CMR imaging with temporary permanent pacemaker protocol is an effective and safe early screening tool for myocardial disease in patients presenting with atrioventricular block requiring immediate, continuous pacing for bradycardia.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.