Objective: Recently, to further enhance the potential in the management of difficult airways, the highly angulated D-Blade was added to the C-MAC system. The purpose of this study was to investigate the laryngoscopic view and intubation parameters using the new C-MAC D-Blade in comparison to the conventional C-MAC video laryngoscope and Macintosh direct laryngoscope in simulated easy and difficult airways.
Methods:We recruited 26 experienced anaesthesia providers into a randomized trial. Each performed tracheal intubation of a Laerdal SimMan ® manikin with each laryngoscope in the following laryngoscopy scenarios: (1) normal airway, (2) cervical spine immobilization, and (3) tongue edema. The intubation times, success rates, number of intubation attempts, laryngoscopic views, and severity of dental compression were recorded.
Results:In all scenarios, video laryngoscopes provided better laryngeal exposure than the ML and appeared to produce less dental pressure. In the cervical spine immobilization scenario, D-Blade caused less dental pressure and showed better Cormack-Lehane (CL) classes than the other devices (p<0.001). There were no differences between video laryngoscopes in success of tracheal intubation (p>0.05). The CMAC provided the most rapid intubation. The rate of failure was 19% with ML. In the tongue edema scenario, the CMAC provided the most rapid and successful intubation (p<0.001). There were no differences between video laryngoscopes in laryngoscopic views according to CL classification and dental pressure (p>0.05). The rate of failure was 46% with the ML and 7% with the D-Blade.
Conclusion:The CMAC D-Blade caused less dental pressure than the conventional C-MAC and ML in the cervical immobilization scenario. The conventional CMAC performed better than the D-Blade and ML in the tongue edema scenario. These two video laryngoscopes may complement each other in various difficult airway situations.
These data provide evidence for the clinical effectiveness of C-MAC videolaryngoscope in managing the unexpected failed intubations in routine anesthesia care. The C-MAC videolaryngoscope is efficient and safe as a primary rescue device in unexpected failed intubations.
Needle tip visualization during ultrasound-guided regional anesthesia (UGRA) is necessary for safety and efficacy. However, disruption of the image of the needle tip driven toward the target is a general problem, especially for beginners. The purpose of this study was to compare performance parameters between using the Sonoplex and Stimuplex D-Plus echogenic needles in a simulated ultrasound-guided interventional task by inexperienced anaesthesia residents. After a standardized training session, 28 anesthesiology residents performed simulated nerve blocks in a beef phantom with each needle. All ultrasound images were digitally stored for analysis. The absolute time the needle tip was in view, total procedure time, and angle of needle insertion were subsequently measured objectively by two single investigators. The procedures that used the Sonoplex echogenic needle had significantly better tip visibility and shorter total procedure time at insertion angles between 42° and 64° relative to the phantom surface. We have demonstrated that inexperienced users who used the Sonoplex echogenic needle were able to complete the procedure more quickly. Needles with improved visibility would be a very useful addition to UGRA for inexperienced users.
Laparoscopy with pneumoperitoneum at pressures of 10 and 13 mm Hg did not alter the TEG values and low intra-abdominal pressure must be used for peritoneal insufflation.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.