The CCF describes an IOP-independent biomechanical property of the cornea that increases with thicker CCT and decreases with greater age. It is moderately strongly associated with CCT and yet explains more of the interindividual variation in GAT IOP than does CCT. Normalized ORA IOP measurements are not associated with CCT.
he early diagnosis of corneal ectasia is of foremost importance in both screening for refractive surgery and the early treatment of keratoconus. Topography or tomography analysis using either videokeratography or optical coherence tomography instruments can help detect alteration in the shape of the cornea such as thinning and increased curvature. However, these instruments cannot measure the mechanical stability, which is thought to be the initiating event of the disease, even before notable changes in corneal morphology take place.1,2 For this reason, there has been increasing interest in developing instruments to measure the in vivo biomechanical properties of the cornea to aid the diagnosis of an ectasia in a "biomechanical" stage, when topography and tomography are nor-T ABSTRACT PURPOSE: To evaluate the ability of a new combined biomechanical index called the Corvis Biomechanical Index (CBI) based on corneal thickness profile and deformation parameters to separate normal from keratoconic patients.METHODS: Six hundred fifty-eight patients (329 eyes in each database) were included in this multicenter retrospective study. Patients from two clinics located on different continents were selected to test the capability of the CBI to separate healthy and keratoconic eyes in more than one ethnic group using the Corvis ST (Oculus Optikgeräte GmbH, Wetzlar, Germany). Logistic regression was employed to determine, based on Database 1 as the development dataset, the optimal combination of parameters to accurately separate normal from keratoconic eyes. The CBI was subsequently independently validated on Database 2.
RESULTS:The CBI included several dynamic corneal response parameters: deformation amplitude ratio at 1 and 2 mm, applanation 1 velocity, standard deviation of deformation amplitude at highest concavity, Ambrósio's Relational Thickness to the horizontal profile, and a novel stiffness parameter. The receiver operating characteristic curve analysis of the training database showed an area under the curve of 0.983. With a cut-off value of 0.5, 98.2% of the cases were correctly classified with 100% specificity and 94.1% sensitivity. In the validation dataset, the same cut-off point correctly classified 98.8% of the cases with 98.4% specificity and 100% sensitivity.
CONCLUSIONS:The CBI was shown to be highly sensitive and specific to separate healthy from keratoconic eyes. The presence of an external validation dataset confirms this finding and suggests the possible use of the CBI in everyday clinical practice to aid in the diagnosis of keratoconus.[J Refract Surg. 2016;32(12):803-810.]
The cornea demonstrates considerable stiffening with age with the behavior closely fitting an exponential power function typical of collagenous tissue. The increase in stiffness could be related to the additional age-related nonenzymatic cross-linking affecting the stromal collagen fibrils.
The TBI generated by the RF/LOOCV provided greater accuracy for detecting ectasia than other techniques. The TBI was sensitive for detecting subclinical (fruste) ectasia among eyes with normal topography in very asymmetric patients. The TBI may also confirm unilateral ectasia, potentially characterizing the inherent ectasia susceptibility of the cornea, which should be the subject of future studies. [J Refract Surg. 2017;33(7):434-443.].
Keratoconic eyes demonstrated less resistance to deformation than normal eyes with similar IOP. The stiffness parameters may be useful in future biomechanical studies as potential biomarkers. [J Refract Surg. 2017;33(4):266-273.].
Strip extensometry tests are usually considered less reliable than trephinate inflation tests in studying corneal biomechanics. In spite of the evident simplicity of strip extensometry tests, several earlier studies preferred inflation tests in determining the constitutive relationship of the cornea and its other material properties, such as Young's modulus and the hysteresis behaviour. In this research, the deficiencies of the strip tests are discussed and a mathematical procedure presented to take account of these deficiencies when obtaining the corneal material properties. The study also involves testing 10 pairs of porcine corneas using both strip extensometry and trephinate inflation techniques and the results are subjected to mathematical back analysis in order to determine the stress-strain behaviour. The behaviour obtained from the strip extensometry tests and using the new mathematical analysis procedure is shown to match closely the inflation test results.
Purpose:
This study aims to introduce and clinically validate a new algorithm that can determine the biomechanical properties of the human cornea
in vivo
.
Methods:
A parametric study was conducted involving representative finite element models of human ocular globes with wide ranges of geometries and material biomechanical behavior. The models were subjected to different levels of intraocular pressure (IOP) and the action of external air puff produced by a non-contact tonometer. Predictions of dynamic corneal response under air pressure were analyzed to develop an algorithm that can predict the cornea's material behavior. The algorithm was assessed using clinical data obtained from 480 healthy participants where its predictions of material behavior were tested against variations in central corneal thickness (CCT), IOP and age, and compared against those obtained in earlier studies on
ex-vivo
human ocular tissue.
Results:
The algorithm produced a material stiffness parameter (Stress-Strain Index or SSI) that showed no significant correlation with both CCT (
p
> 0.05) and IOP (
p
> 0.05), but was significantly correlated with age (
p
< 0.01). The stiffness estimates and their variation with age were also significantly correlated (
p
< 0.01) with stiffness estimates obtained earlier in studies on
ex-vivo
human tissue.
Conclusions:
The study introduced and validated a new method for estimating the
in vivo
biomechanical behavior of healthy corneal tissue. The method can aid optimization of procedures that interfere mechanically with the cornea such as refractive surgeries and introduction of corneal implants.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.