Energy storage and return feet are designed for active amputees. However, little is known about the socket reaction moments in transtibial prostheses with energy storage and return feet. The aim of this study was to investigate the effect of alignment changes on the socket reaction moments during gait while using the energy storage and return feet.
Methods
A Smart Pyramid™ was used to measure the socket reaction moments in 10 subjects with transtibial prostheses while walking under 25 alignment conditions, including a nominal alignment (as defined by conventional clinical methods), as well as angle malalignments of 2°, 4° and 6° (flexion, extension, abduction, and adduction) and translation malalignments of 5mm, 10mm and 15mm (anterior, posterior, lateral, and medial) referenced from the nominal alignment. The socket reaction moments of the nominal alignment were compared with each malalignment.
Findings
Both coronal and sagittal alignment changes demonstrated systematic effects on the socket reaction moments. In the sagittal plane, angle and translation alignment changes demonstrated significant differences (P<0.05) in the minimum moment, the moment at 45% of stance and the maximum moment for some comparisons. In the coronal plane, angle and translation alignment changes demonstrated significant differences (P<0.05) in the moment at 30% and 75% of stance for all comparisons.
Interpretation
The alignment may have systematic effects on the socket reaction moments in transtibial prostheses with energy storage and return feet. The socket reaction moments could potentially be a useful biomechanical parameter to evaluate the alignment of the transtibial prostheses.
The alignment of a lower limb prosthesis affects the way load is transferred to the residual limb through the socket, and this load is critically important for the comfort and function of the prosthesis. Both magnitude and duration of the moment are important factors that may affect the residual limb health. Moment impulse is a well-accepted measurement that incorporates both factors via moment–time integrals. The aim of this study was to investigate the effect of alignment changes on the socket reaction moment impulse in transtibial prostheses. Ten amputees with transtibial prostheses participated in this study. The socket reaction moment impulse was measured at a self-selected walking speed using a Smart Pyramid™ in 25 alignment conditions, including a nominal alignment (clinically aligned by a prosthetist), as well as angle malalignments of 2°, 4° and 6° (abduction, adduction, extension and flexion) and translation malalignments of 5 mm, 10 mm and 15 mm (lateral, medial, anterior and posterior). The socket reaction moment impulse of the nominal alignment was compared for each condition. The relationship between the alignment and the socket reaction moment impulse was clearly observed in the coronal angle, coronal translation and sagittal translation alignment changes. However, this relationship was not evident in the sagittal angle alignment changes. The results of this study suggested that the socket reaction moment impulse could potentially serve as a valuable parameter to assist the alignment tuning process for transtibial prostheses. Further study is needed to investigate the influence of the socket reaction moment impulse on the residual limb health.
Abstract-A person with amputation's subjective perception is the only tool available to describe fit and comfort to a prosthetist. However, few studies have investigated the effect of alignment on this perception. The aim of this article is to determine whether people with amputation could perceive the alignment perturbations of their prostheses and effectively communicate them. A randomized controlled perturbation of angular (3 and 6 degrees) and translational (5 and 10 mm) alignments in the sagittal (flexion, extension, and anterior and posterior translations) and coronal (abduction, adduction, and medial and lateral translations) planes were induced from an aligned condition in 11 subjects with transtibial prostheses. The perception was evaluated when standing (static) and immediately after walking (dynamic) using software that used a visual analog scale under each alignment condition. In the coronal plane, Friedman test demonstrated general statistical differences in static (p < 0.001) and dynamic (p < 0.001) measures of perceptions with angular perturbations. In the sagittal plane, it also demonstrated general statistical differences in late-stance dynamic measures of perceptions (p < 0.001) with angular perturbations, as well as in early-stance dynamic measures of perceptions (p < 0.05) with translational perturbations. Fisher exact test suggested that people with amputation's perceptions were good indicators for coronal angle malalignments but less reliable when defining other alignment conditions.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.