BackgroundHigh costs still limits the widespread use of xenon in the clinical practice. Therefore, we evaluated xenon consumption of different delivery modes during general surgery.MethodsA total of 48 patients that underwent general surgery with balanced xenon anaesthesia were retrospectively analysed according to the mode of xenon delivery during maintenance phase (ECO mode, AUTO mode or MANUAL mode).ResultsXenon consumption was highest during the wash-in phase (9.4 ± 2.1l) and further decreased throughout maintenance of anaesthesia. Comparison of different xenon delivery modes revealed significant reduced xenon consumption during ECO mode (18.5 ± 3.7L (ECO) vs. 24.7 ± 11.5L (AUTO) vs. 29.6 ± 14.3L (MANUAL); p = 0.033). No differences could be detected with regard to anaesthetic depth, oxygenation or performance of anaesthesia.ConclusionThe closed-circuit respirator Felix Dual offers effective reduction of xenon consumption during general surgery when ECO mode is used.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.