A rapid and sensitive RP-HPLC method with UV detection (262 nm) for routine analysis of adefovir dipivoxil in bulk and in pharmaceutical formulation was developed. Chromatography was performed with mobile phase containing a mixture of acetonitrile and phosphate buffer (50:50, v/v) with flow rate 1.0 mL min-l. In the range of 5.0-100 µg/mL, the linearity of adefovir dipivoxil shows a correlation co-efficient of 0.9999. The proposed method was validated by determining sensitivity accuracy, precision, robustness stability, specificity, selectivity and system suitability parameters.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.