Among the possible variables responsible for systemic errors in the acquisition of pachometric measurements, the corneal hydration condition plays a determinant role. In the present study, the authors have investigated the importance of this condition with two different pachometers (Storz Corneoscan II and Allergan Humphrey), reproducing an experimental dehydrated condition. The main datum arising from the comparison of the measurements obtained is represented by the different time course of the value of the measurement in the two groups. In fact, confronting the medium initial superimposed thickness value in the two tables (521 vs. 520 µm) there is a substantial difference in the final examination. For the first group of 60 corneas, the thickness values appear almost identical in the different intervals of time (medium initial thickness 521 vs. 512 µm final thickness). For the group examined with the solid-point machine, a decreasing thickness equal to 7% was evidenced (520 vs. 482 µm) with a significant decrease in the corneal thickness after only 45 s.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.