A pulsed distributed feedback quantum cascade laser operating near 970 cm -1 (10.3 µm) was coupled with the technique of cavity ring-down spectroscopy, as described here for the first time. The newly constructed set-up was tested by recording three relatively weak rotational lines of the 10 0 0→00 0 1 vibrational band of CO 2 in the range from 966.75 cm -1 to 971.5 cm -1 . The CO 2 lines were recorded by measuring the decay time of a CO 2 -N 2 mixture flowing through an open sample tube placed between the cavity ring-down mirrors. The quantum cascade laser frequency was tuned at a rate of ~ 0.071 cm -1 /K by changing the heat sink temperature in the range between -20 and 50 °C. The first results demonstrated the applicability and high sensitivity of the cavity ring-down spectroscopy -pulsed quantum cascade laser combination and encouraged us to extend our research to the study and detection of ammonia. We demonstrated that a detection limit of ammonia of ~ 25 ppbv can be attained with the current set-up. Basic instrument performance and optimization of the experimental parameters for sensitivity improvement are discussed.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.