New mathematical algorithms have been applied to a computer controlled closed breathing circuit system for non-invasive measurement of cardiac output (COniv). This system has been described in an animal study. Forty patients were studied 5 and 18 hours after cardiac surgery using the thermodilution technique as the reference (COtd). The variables entered into the algorithms for COniv were oxygen uptake, carbon dioxide elimination, end-tidal carbon dioxide partial pressure, tidal volume and arterial oxygen saturation. Mixed venous carbon dioxide partial pressure was obtained from an automatically implemented short rebreathing manoeuvre. Pulmonary perfusion was calculated by a modified Fick equation for carbon dioxide and the shunt flow added to obtain COniv. During mechanical ventilation, there was a good agreement between COtd and COniv (r = 0.8). The bias was -0.14 l/min and the precision was 0.77 l/min. The reproducibility of COniv was 0.03 l/min and for COtd -0.03 l/min with a standard deviation of the difference being 0.35 l/min for COniv and 0.31 l/min for COtd. In awake, but sedated extubated patients, the method proved unsatisfactory on account for uneven tidal volumes and difficulties with leakage around the mouth piece. We conclude that this new technique provides reliable and reproducible measures of cardiac output in sedated, ventilated patients.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.