Results of a multicenter evaluation of automated assays for measurement of apolipoproteins (apo) A-I and B with the Paramax analytical system are reported. Apo A-I and apo B response surface models were used to optimize concentrations of critical assay variables. Overall imprecision for apo A-I controls at concentrations of 1.01-1.61 g/L was 3.7-6.6%; overall imprecision for apo B controls at 1.00-1.61 g/L was 2.8-6.9%. There was no interference in apo A-I measurements. Albumin concentrations > 59 g/L resulted in a negative interference, and collection in sodium heparin caused a positive interference in apo B results. Apo A-I and apo B assays demonstrated acceptable agreement with comparative methods, although the Paramax apo B assays had a negative bias with respect to comparison methods. In 116 healthy individuals, serum apo A-I ranged from 0.97 to 2.05 g/L and serum apo B ranged from 0.51 to 1.32 g/L.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.