Влияние Состояния Рабочей Поверхности Гальванического Датчика Импульсного Рентгеновского Излучения На Его Быстродействие

Abstract: A new design of a galvanic sensor of pulsed X-ray radiation is proposed, which is a flat electric capacitor with a window in one metal lining and a solid dielectric made of the single crystal sapphire with a thickness of 200-300 µm inside. The influence of the surface roughness of the sapphire in the window area on the galvanic linear sensor of X-ray radiation has been established. Tests have shown that with ultra-smooth polishing of the sapphire plate working surface in the window area to a roughness of Rq≤ 0… Show more