Validation of the Finite Element Model versus Biomechanical Assessments of Dental Implants and Total Knee Replacements

Kwan-Su Kang,
Kwang-Min Park,
Jin-Woo Ahn
et al.

Abstract: Computer modeling and simulation (CM&S) technology is widely used in the medical device industry due to its advantages such as reducing testing time and costs. However, the developer’s parameter settings during the modeling and simulation process can have a significant impact on the results. This study developed a test model for the rotational shear strength of dental implants and the constraint force of total knee replacements based on CM&S technology and proposes ideal parameters to ensure reliabilit… Show more