Validation of temporal gait metrics from three IMU locations to the gold standard force plate

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
1
1

Citation Types

4
11
0

Year Published

2017
2017
2024
2024

Publication Types

Select...
7
1
1

Relationship

0
9

Authors

Journals

citations
Cited by 13 publications
(15 citation statements)
references
References 34 publications
4
11
0
Order By: Relevance
“…The ankle-positioned IMUs were valid for the assessment of the capacity-based number of steps, walking time, gait speed, and stride length when tested during a standardized measurement protocol, including turns, slaloms, and short walking bouts. The validity of the IMU-based assessment of capacity-based gait speed, stride length, and number of steps during regular walking is in line with previous studies using ankle-worn sensors [ 29 , 33 , 34 , 35 ]. A limitation of the algorithm was that one sensor sometimes missed stride detection, which occurred in 8% during standardized testing and in 16% beyond standardized testing.…”
Section: Discussionsupporting
confidence: 85%
See 1 more Smart Citation
“…The ankle-positioned IMUs were valid for the assessment of the capacity-based number of steps, walking time, gait speed, and stride length when tested during a standardized measurement protocol, including turns, slaloms, and short walking bouts. The validity of the IMU-based assessment of capacity-based gait speed, stride length, and number of steps during regular walking is in line with previous studies using ankle-worn sensors [ 29 , 33 , 34 , 35 ]. A limitation of the algorithm was that one sensor sometimes missed stride detection, which occurred in 8% during standardized testing and in 16% beyond standardized testing.…”
Section: Discussionsupporting
confidence: 85%
“…However, possible (slight) deviations from actual therapy time were not registered for reasons of feasibility and privacy. Although the IMU-based gait algorithms have shown to be valid [ 25 , 34 , 35 ], it still remains challenging to validly measure impaired gait. We, therefore, validated the gait algorithm during the gait capacity assessment.…”
Section: Discussionmentioning
confidence: 99%
“…These datasets have then been used to validate the methods. One study has presented the validation of temporal gait metrics from three IMU locations to the GA using Bland-Altman diagrams [25,26]. We have chosen to use a structured methodology that adopts a robust approach to complete the relative task for the validation protocol, as required by the general methodology defined in [24], to validate the use of the new system.…”
Section: Introductionmentioning
confidence: 99%
“…Some works exist regarding validation analysis of wearable systems for gait analysis based on processing of kinematic signals. These studies address comparative analyses with clinical instruments, such as instrumented treadmill [ 49 ], force platform [ 50 ] or pressure sensitive walkway (GAITRite ) [ 35 , 51 , 52 ]. No works presenting a comparative analysis with the gold standard (stereophotogrammetry system) has been found.…”
Section: Discussionmentioning
confidence: 99%