“…As no major deviations of the DF measured from the DF targeted could be measured (S1 the dilution factors, as is standard practice in other works as well [5,11,16,19]. The intercept, slope and R 2 of the straight-line fit of both methods are reported in Table 2.…”
Section: Results Related To Total Cell Densitymentioning
confidence: 99%
See 2 more Smart Citations
“…As no major deviations of the DF measured from the DF targeted could be measured (S1 the dilution factors, as is standard practice in other works as well [5,11,16,19]. The intercept, slope and R 2 of the straight-line fit of both methods are reported in Table 2.…”
Section: Results Related To Total Cell Densitymentioning
confidence: 99%
“…The measured TCDs of set A were plotted against the DF targeted for both methods in Fig 2 and a straight-line fit was applied. At this point, we would like to emphasize that the plotted measured TCDs do not represent the raw TCDs measured in reality, as they have been extrapolated using the dilution factors, as is standard practice in other works as well [5, 11, 16, 19]. The intercept, slope and R 2 of the straight-line fit of both methods are reported in Table 2.…”
Section: Resultsmentioning
confidence: 99%
See 1 more Smart Citation
“…In this context, cell viability and cell numbers per target dose can be used as product purity parameters and associated potency based on the assumed mechanism of action 26 to monitor ATMP-manufacturing processes and is also often selected as one of the IPCs for critical steps. Various analytical methods are used for cell counting and viability, including manual trypan blue exclusion, 27 automatic cell counters, 28 , 29 flow cytometry-based methods using live/dead cell markers or Trucount beads, 30 and built-in fluidic-based system based on volumetric method that enable precise automated cell counting in a single-platform flow cytometry. 31 However, real-time analysis of cellular products remains difficult, especially for adherent cells.…”
Section: The Abc Of Analytical Qc Methods For Atmp Manufacturingmentioning
confidence: 99%