2012
DOI: 10.1128/cvi.00023-12
|View full text |Cite
|
Sign up to set email alerts
|

Validation of a Single-Platform, Volumetric, CD45-Assisted PanLeucogating Auto40 Flow Cytometer To Determine the Absolute Number and Percentages of CD4 T Cells in Resource-Constrained Settings Using Cameroonian Patients' Samples

Abstract: ABSTRACTThe study evaluated the single-platform, volumetric, CD45-assisted PanLeucogating Auto40 flow cytometer (Apogee Flow Systems Ltd., Hemel Hempstead, United Kingdom) for CD4 T cell numeration, compared to the reference FACSCalibur flow cytometer. Results of absolute counts and percentages of CD4 T cells by Auto40 and FACSCalibur of 234 tripotassium EDTA (K3-EDTA)-blood samples from 146 adults and 88 children (aged from 18 months to 5 years), living in Yaoundé, Cameroon, w… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
2

Citation Types

2
14
0

Year Published

2012
2012
2020
2020

Publication Types

Select...
6

Relationship

2
4

Authors

Journals

citations
Cited by 10 publications
(16 citation statements)
references
References 30 publications
2
14
0
Order By: Relevance
“…Higher precisions for high than for low CD4 T cell count have been previously reported with single-platform volumetric flow cytometric method [40]. Finally, the Auto40 system showed a high capacity of accurately discriminate CD4 T cell values around the relevant CD4 T cells thresholds recommended by the 2010-revised WHO guidelines to initiate and follow ART in HIV-infected adults or children, as previously reported [15,19,46]. …”
Section: Discussionsupporting
confidence: 74%
See 3 more Smart Citations
“…Higher precisions for high than for low CD4 T cell count have been previously reported with single-platform volumetric flow cytometric method [40]. Finally, the Auto40 system showed a high capacity of accurately discriminate CD4 T cell values around the relevant CD4 T cells thresholds recommended by the 2010-revised WHO guidelines to initiate and follow ART in HIV-infected adults or children, as previously reported [15,19,46]. …”
Section: Discussionsupporting
confidence: 74%
“…In the present study, we carried out a new evaluation of the biological performances of the Auto40 flow cytometer in Chad focusing on the HIV-infected adult and children populations, independently of the manufacturer, and in complement to those previously reported in Senegal [15] and Cameroon [19]. Indeed, several evaluations by at least 3 independent laboratories in the field are strongly recommended for effective validation of the newly introduced affordable CD4 T cell measurement technologies to be used in resources-constrained settings [6,21].…”
Section: Discussionmentioning
confidence: 99%
See 2 more Smart Citations
“…Moreover, they require highly trained personnel, an air conditioned lab and a cold chain to ship and store reagents, facilities which are not always available in resource-limited settings. To ensure decentralization of the HIV-monitoring services, more affordable CD4 dedicated flow cytometers have been introduced on the market, including FACSCount (BD Biosciences), CyFlow Counter (Partec), Guava EasyCD4 (Millipore), Apogee Auto40 (Apogee), PointCare NOW (PointCare) [ 5 13 ]. At the end of the previous decade, low-cost Point of Care (POC) technologies were being developed, mainly based on digital imaging [ 14 18 ].…”
Section: Introductionmentioning
confidence: 99%