2017
DOI: 10.1002/acm2.12025
|View full text |Cite
|
Sign up to set email alerts
|

Validation of a modern second‐check dosimetry system using a novel verification phantom

Abstract: PurposeTo evaluate the Mobius second‐check dosimetry system by comparing it to ionization‐chamber dose measurements collected in the recently released Mobius Verification Phantom™ (MVP). For reference, a comparison of these measurements to dose calculated in the primary treatment planning system (TPS), Varian Eclipse with the AcurosXB dose algorithm, is also provided. Finally, patient dose calculated in Mobius is compared directly to Eclipse to demonstrate typical expected results during clinical use of the Mo… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
4
1

Citation Types

0
7
0

Year Published

2019
2019
2023
2023

Publication Types

Select...
9
1

Relationship

0
10

Authors

Journals

citations
Cited by 34 publications
(7 citation statements)
references
References 24 publications
0
7
0
Order By: Relevance
“…Reports and scientific publications that describe the accuracy of independent IMRT dose or monitor unit calculation software, whether developed in‐house or commercial products, are scarce. Most publications deal with static (step‐and shoot) or dynamic IMRT delivery based on a multileaf collimator or VMAT, 43‐45 and only a few of them with advanced treatment techniques performed with a robotic linac 46 and TomoTherapy 47 . Nevertheless, the commercial vendor list (Table 2) reveals that these tools are clinically applied for verifying IMRT or VMAT.…”
Section: Role Of Dose/mu Verifications In a Comprehensive Qa Programmentioning
confidence: 99%
“…Reports and scientific publications that describe the accuracy of independent IMRT dose or monitor unit calculation software, whether developed in‐house or commercial products, are scarce. Most publications deal with static (step‐and shoot) or dynamic IMRT delivery based on a multileaf collimator or VMAT, 43‐45 and only a few of them with advanced treatment techniques performed with a robotic linac 46 and TomoTherapy 47 . Nevertheless, the commercial vendor list (Table 2) reveals that these tools are clinically applied for verifying IMRT or VMAT.…”
Section: Role Of Dose/mu Verifications In a Comprehensive Qa Programmentioning
confidence: 99%
“…[8][9][10] Many studies have evaluated the feasibility of a log-based PSQA for tomotherapy. 8,9,[11][12][13] However, for the tomotherapy PSQA, since no log-based commercial software is currently available, the cylindrical Virtual Water™ phantom (Gammex RMI, Middleton, WI; called the "Cheese" phantom), which is available upon the purchase of a tomotherapy machine, is widely used.…”
Section: Introductionmentioning
confidence: 99%
“…It performs a full recalculation of dose on the patient CT and allows for quality assurance of the treatment plan by offering a “delivered dose” calculation, generated using the M3D model and treatment machine's log files. 18 , 19 …”
Section: Introductionmentioning
confidence: 99%