2015
DOI: 10.1016/j.hrthm.2015.07.013
|View full text |Cite
|
Sign up to set email alerts
|

User-reported abrasion-related lead failure is more common with durata compared to other implantable cardiac defibrillator leads

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
4
1

Citation Types

0
7
0

Year Published

2017
2017
2023
2023

Publication Types

Select...
8

Relationship

0
8

Authors

Journals

citations
Cited by 17 publications
(8 citation statements)
references
References 11 publications
(7 reference statements)
0
7
0
Order By: Relevance
“…However, the Optim layer does not cover the silicone elastomer insulation under the shock coils. Recent reports have described insulation failure of the Durata lead 6, 7, 8, 9. To prevent tissue ingrowth, the cross section of the outer wire of the shock coil is not round but flat and thin.…”
Section: Discussionmentioning
confidence: 99%
“…However, the Optim layer does not cover the silicone elastomer insulation under the shock coils. Recent reports have described insulation failure of the Durata lead 6, 7, 8, 9. To prevent tissue ingrowth, the cross section of the outer wire of the shock coil is not round but flat and thin.…”
Section: Discussionmentioning
confidence: 99%
“…Could the Optim insulation itself be the problem? Although failure and abrasions have been reported on the Riata ST Optim and Durata defibrillator leads, prospective data on these leads seems to be reliable . On the other hand, the Optim insulation for the Tendril leads probably is thinner than and not as strong as that for the Durata leads.…”
Section: Discussionmentioning
confidence: 91%
“…However, previous single case reports show failure of the HV circuitry and failure to defibrillate despite nominal low‐voltage shock impedance measurements 4,10 . Analyzing both internal and external insulation breaches, Shah et al reported that three of four patients with Durata ICD leads (St. Jude Medical, St. Paul, MN) presented with failure of HV shock therapy or death without a history of electrical malfunction 11 . With the goal to assess internal insulation breaches in Durata ICD leads, Hauser et al searched the Food and Drug Administration Manufacturer and User Facility Device Experience database 12 .…”
Section: Discussionmentioning
confidence: 99%