Abstract: Implantation of new large cobalt-chromium stents is a good therapeutic option for the treatment of native and recurrent CoA.

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
2

Citation Types

0
9
1

Year Published

2017
2017
2024
2024

Publication Types

Select...
7

Relationship

2
5

Authors

Journals

citations
Cited by 11 publications
(10 citation statements)
references
References 14 publications
0
9
1
Order By: Relevance
“…The result of this study is in contrast with the results presented by Chessa et al [7] and Alcibar et al [8]. In our group of patients, similarly to others [6,[9][10][11], no deaths were observed during or after SI. Data of surgical treatment of aortic coarctation are confounded because of heterogeneous subject population, incomplete follow-up, and frequently unclear distinctions between planned and unplanned intervention.…”
Section: Discussioncontrasting
confidence: 99%
See 2 more Smart Citations
“…The result of this study is in contrast with the results presented by Chessa et al [7] and Alcibar et al [8]. In our group of patients, similarly to others [6,[9][10][11], no deaths were observed during or after SI. Data of surgical treatment of aortic coarctation are confounded because of heterogeneous subject population, incomplete follow-up, and frequently unclear distinctions between planned and unplanned intervention.…”
Section: Discussioncontrasting
confidence: 99%
“…According to the published data of Meadows et al [11], CP stent fractures were noted in two patients after one year, and in 11 patients after two years (among 104 treated), with evidence of fracture progression. From our experience (presented elsewhere [6]) we observed no fractures in any of the 48 Andrastents XL and XXL that we used to dilate native coarctation of the aorta and ReCoA.…”
Section: Discussionmentioning
confidence: 72%
See 1 more Smart Citation
“…However, one must note that stent fracture and progression of the fracture can occur when Cheatham Platinum stents are applied [10]. In contrast, in our recently published series of Andrastents used in CoA this complication was not observed [6]. …”
Section: Discussionmentioning
confidence: 99%
“…Different stents were applied in this period depending on availability (bare metal: Palmaz, Cheatham Platinum (CP), Andrastents XL/XXL and covered with polytetrafluoroethylene CP stents). Characteristics of those stents have been described elsewhere [5, 6]. The stents were implanted retrogradely with the standard technique.…”