Use of high resolution/accurate mass full scan/data-dependent acquisition for targeted/non-targeted screening in equine doping control

Abstract: High-resolution mass spectrometry (HRMS) is a very powerful technology for equine doping control analysis. The more recently developed hybrid type of Orbitrap-based HRMS instrument allows for both targeted and non-targeted...

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
2
1

Citation Types

1
7
0

Year Published

2021
2021
2024
2024

Publication Types

Select...
6

Relationship

1
5

Authors

Journals

citations
Cited by 8 publications
(8 citation statements)
references
References 31 publications
1
7
0
Order By: Relevance
“…Stock ammonium formate (HCOONH 4 ) buffer solution comprising 1.0 mol/L (M) HCOONH 4 and 1.0 M FA (pH 3.6) and working buffer solution containing 5 mM each of HCOONH 4 and FA were made. The details about the drug standards, solvents, chemicals, reagents, and their preparations are described elsewhere . The name, chemical formula, exact mass of the protonated species, RT, and initial concentration of the 57 drugs spiked into equine plasma are listed in Table S1.…”
Section: Methodsmentioning
confidence: 99%
See 4 more Smart Citations
“…Stock ammonium formate (HCOONH 4 ) buffer solution comprising 1.0 mol/L (M) HCOONH 4 and 1.0 M FA (pH 3.6) and working buffer solution containing 5 mM each of HCOONH 4 and FA were made. The details about the drug standards, solvents, chemicals, reagents, and their preparations are described elsewhere . The name, chemical formula, exact mass of the protonated species, RT, and initial concentration of the 57 drugs spiked into equine plasma are listed in Table S1.…”
Section: Methodsmentioning
confidence: 99%
“…This LOD is close to that achieved using the targeted analytical method. 21 In addition, 26 of the 55 drugs had a low LOD (0.05 to 0.2 ng/mL), 16 of the drugs had a slightly higher LOD (0.5 to 2 ng/mL), and 5 drugs (piroxicam, 2-(1-hydroxyethyl)promazine sulfoxide, benzoylecgonine, theophylline, and caffeine) had a relatively high LOD (5 to 100 ng/mL). Although the LOD for caffeine was 100 ng/mL, it was actually detected at 0.1 ng/mL by the CD processing workflow but was filtered out by ROM (Table S6).…”
Section: Untargeted Detection Of Drugs Spiked Into Blankmentioning
confidence: 98%
See 3 more Smart Citations