1998
DOI: 10.1002/(sici)1521-4168(19980801)21:8<450::aid-jhrc450>3.0.co;2-j
|View full text |Cite
|
Sign up to set email alerts
|

Use of a Drying Cartridge in On-Line Solid-Phase Extraction-Gas Chromatography-Mass Spectrometry

Abstract: A drying cartridge was used and optimized for the in‐line elimination of water from the desorption eluent in on‐line solid phase extraction–gas chromatography (SPE–GC). The cartridge is essentially a small stainless‐steel precolumn packed with a drying agent which can be regenerated by simultaneous heating and purging with a moisture‐free gas. The drying cartridge was mounted on an additional valve instead of between the SPE–GC transfer valve and the on‐column injector to enable regeneration of the cartridge d… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
1
1

Citation Types

0
8
0

Year Published

2003
2003
2011
2011

Publication Types

Select...
4
1

Relationship

0
5

Authors

Journals

citations
Cited by 23 publications
(8 citation statements)
references
References 16 publications
0
8
0
Order By: Relevance
“…Since water is not a good solvent for GC, primarily because it hydrolyses the siloxane bonds in GC columns causing deterioration of the column performance, the introduction of water directly to GC should be avoided. After trapping, and before elution of the analytes, the SPE column is often dried with a gas flow, or the extract is dried with a separate drying column packed with copper sulphate or silica to remove water, which is placed after the SPE column [5,6,9,10]. Of these techniques, the use of a nitrogen purge is the more straightforward and easy-to-use.…”
Section: Solid-phase Extractionmentioning
confidence: 99%
See 1 more Smart Citation
“…Since water is not a good solvent for GC, primarily because it hydrolyses the siloxane bonds in GC columns causing deterioration of the column performance, the introduction of water directly to GC should be avoided. After trapping, and before elution of the analytes, the SPE column is often dried with a gas flow, or the extract is dried with a separate drying column packed with copper sulphate or silica to remove water, which is placed after the SPE column [5,6,9,10]. Of these techniques, the use of a nitrogen purge is the more straightforward and easy-to-use.…”
Section: Solid-phase Extractionmentioning
confidence: 99%
“…Methods based on both SPE-GC and LLE-GC have been developed for a variety of environmental determinations, and SPE-GC has also been used for the analysis of drugs in biological fluids [4,5,6,7,8,9,10,11,12,13,14,15,16,17,18,19,20,21,22,23,24,25,26,27,28,29,30,31,32,33,34,35,36,37,38,39,40,41,42,43,44,53,54,55,56,57,58,59,60,61,62,…”
Section: Liquid Samplesmentioning
confidence: 99%
“…It is used for matrix separation, rapid clean-up and enrichment of target compounds preceding chromatographic analysis [2][3][4][5]. The technical principle is based on the distribution of analytes between a solid and a liquid or a headspace vapour.…”
Section: Introductionmentioning
confidence: 99%
“…1) [3,4,6]. Many of the reported methods reduce the remaining water volume on the phase material after the extraction step by actively drying the sorbent [4,5,[7][8][9][10][11][12]. The reasons for this step are manifold [13,14].…”
Section: Introductionmentioning
confidence: 99%