2014
DOI: 10.1016/j.hrthm.2014.04.012
|View full text |Cite
|
Sign up to set email alerts
|

Use of a discrimination algorithm to reduce inappropriate shocks with a subcutaneous implantable cardioverter-defibrillator

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
2
1

Citation Types

0
54
0

Year Published

2016
2016
2021
2021

Publication Types

Select...
5
4

Relationship

0
9

Authors

Journals

citations
Cited by 85 publications
(56 citation statements)
references
References 22 publications
0
54
0
Order By: Relevance
“…This significantly reduces rates of IAS 47. In the IDE study, dual zone programming reduced the 2-year IAS rate from 26.4% to 10.3% 48. Consequently, Burke et al reported a 34% decrease in 6-month incidence of IAS from the first quartile of patients enrolled into their combined registry compared with the last 6.…”
Section: Discussionmentioning
confidence: 99%
“…This significantly reduces rates of IAS 47. In the IDE study, dual zone programming reduced the 2-year IAS rate from 26.4% to 10.3% 48. Consequently, Burke et al reported a 34% decrease in 6-month incidence of IAS from the first quartile of patients enrolled into their combined registry compared with the last 6.…”
Section: Discussionmentioning
confidence: 99%
“…rate, onset, stability, and atrioventricular relationship), after which morphology-based discrimination is applied, the S-ICD system initially analyzes the morphological characteristics of the electrogram recorded by the device to determine the presence of an arrhythmia. 1 The S-ICD system evaluates up to 41 fiduciary points of each ventricular complex in order to accurately discriminate VT from SVT. The sensitivity of ventricular arrhythmia detection by S-ICD has been shown to be equivalent to transvenous ICD systems, and the specificity for SVT may be significantly better for the S-ICD system, especially when dual-zone programming (with a conditional zone) is applied.…”
Section: Discussionmentioning
confidence: 99%
“…In addition, beat-to-beat morphology variations are also examined and, based on these three parameters, the S-ICD determines whether therapy should be delivered. 36 Confirmation of a shockable rhythm is ensured separately prior to capacitor charging and shock delivery. Preimplant ECG screening with patients in different postures is crucial to identifying patients unsuited for S-ICD use, further lowering the rate of inappropriate therapy.…”
Section: Subcutaneous Implantable Cardioverter-defibrillatorsmentioning
confidence: 99%